background image

 

Spirobank II

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rev.1.6 

Page 5 of 35 

User manual 

1.1.4

 

Patient effect on the use of the device 

 

A spirometry test should only be carried out when the patient is at rest and in good health, in suitable testing conditions. A spirometry 
test requires the full 

collaboration

 of the patient since she/he must perform a complete forced expiration, in order to obtain a reliable 

test result. 

 

 

1.1.5

 

Limitations of use 

 Contraindications 

 

An analysis of the results of a spirometry test is not by itself sufficient to make a correct diagnosis of the patient’s clin

ical condition. A 

detailed clinical history of the patient is also required together with the results of any other test(s) suggested by a doctor. 

Test comments, a test interpretation and suggested therapeutic treatment must be given by a doctor. 
Any symptoms that the patient has at the time of the test must be carefully considered before a spirometry test is made. The user is 
responsible to assess both the mental and the physical condition of the patient in order to perform a proper test, furthermore, in the 
evaluation of test results, the user must also assess the degree of collaboration of each test carried out. 

A spirometry test requires the full collaboration of the patient. The results depend on the person’s ability to inspire as mu

ch air as possible 

and to expire all of the air as fast and for as long as possible. If these fundamental conditions are not respected then the results obtained 

during spirometry testing will not be considered accurate, and therefore the test results are “not acceptable”.

 

The acceptability of a test is the responsibility of the doctor. Special attention should be given when testing elderly patients, children and 

handicapped people. 
The device should not be used if any conceivable or actual anomalies or malfunctions appear which may compromise the accuracy of 
the results. 
 

 

WARNING 

When used as a pulse oximeter the Spirobank II has limited alarms, therefore the device requires frequent display observation 
of SpO2 and pulse rate. 

 

1.2

 

Important safety warnings 

 

Spirobank II

 has been examined by an independent laboratory which has certified the compliance of the device to the European 

Safety Standards EN 60601-1 and guarantees the EMC Requirements within the limits laid down in the European Standard EN 60601-
1-2. 

Spirobank II

 is continuously checked during manufacturing and therefore the product complies with the established security levels and 

quality standards laid down by the Council Directive 93/42/EEC for MEDICAL DEVICES. 
After removing the device from its packaging, check to see that there is no visible damage. In case of damage do not use the device and 
return it to the manufacturer for repair. 

 

 

WARNING  

The safety and the correct performance of the device can only be assured if the user respects all of the relevant safety rules 
and regulations.  

The manufacturer will not be held responsible for damage due to user’s neglect to correctly to follow these ins

tructions.  

The  device  must  be  used  only  and  exclusively  as  a  spirometer  following  the  indications  given  by  the  manufacturer  with 
particular attention to the paragraph on INTENDED USE, and utilizing only original spare parts and accessories. Use of 
non-original parts such as the turbine flow sensor and oximetry sensor or other accessories may cause errors in measurement 
and/or compromise the correct functioning of the device, and is therefore not permitted. 

In  particular,  the  use  of  cables  other  than  those  specified  by  the  manufacturer  could  cause  increased  emissions  or  lower 
electromagnetic immunity from the device and result in improper operation. 
The device should not be used beyond the declared life span.. In normal conditions the lifespan of the device is estimated to 
be around 10 years.  

The  device  constantly  monitors  the  state  of  charge  of  this  battery  and  a  message  informs  the  user  when  the  battery  is 
discharged.  
In the event of any incident or accident of any kind resulting from the use of the device, the user is required to inform the 
manufacturer  without  delay,  this  procedure  is  laid  down  in  Article.9  of  the  European  Regulations  No.  46/1997,  which 
implemented the EC Directive No. 93/42. 

 

1.2.1

 

Danger of cross-contamination 

 

Two different types of turbine sensors can be used with the device, one is reusable and the other is single-patient disposable. A disposable 
mouthpiece is required in order to connect a patient to the spirometer. In order to avoid exposing the patient to the critical danger of 
cross-contamination, the reusable flow sensor must always be cleaned before each spirometry test, and a new disposable mouthpiece 
must always be used for each patient. The use of an anti-bacterial filter is at the discretion of the doctor. If a single-patient disposable 

turbine is used, then a new one must be used for each patient. 
 

Содержание Spirobank II

Страница 1: ...Spirobank II Rev 1 6 Page 1 of 35 User manual Spirobank II User manual rev 1 6 Issue date 20 01 2021 Approval date 20 01 2021...

Страница 2: ...HE Spirobank II 17 2 1 Switch on and switch off the device 17 2 2 Energy saving 18 2 3 Main screen 18 2 4 Symbols and Icons 18 2 5 Service menu 19 2 5 1 Reusable turbine calibration 22 2 6 Patient Dat...

Страница 3: ...al packaging In the unlikely event that you have a problem with your device please use the original packaging and return it to the distributor or manufacturer Should this be the case please follow the...

Страница 4: ...he patient by the doctor WARNING The manufacturer cannot be held responsible for any damage caused by the user of the device failing to follow instructions and warnings in this manual If the user of t...

Страница 5: ...Spirobank II is continuously checked during manufacturing and therefore the product complies with the established security levels and quality standards laid down by the Council Directive 93 42 EEC for...

Страница 6: ...spirometer are only to be used as a reference guide to purchase the correct size mouthpiece required These mouthpieces are clean but not sterile To purchase appropriate mouthpieces generally either pa...

Страница 7: ...vent safety hazards for the user 1 2 5 USB connection cable Incorrect use or application of the USB cable may produce inaccurate measurements wich will show very inaccurate values of the patient s con...

Страница 8: ...may be damaged or suffer a drop in overall performance The internal battery pack safety sensor could also be damaged as well by any of the above events Furthermore the user of the device could be har...

Страница 9: ...device or the battery charger and do not use it any longer as any of these events may cause acid leakage overheating smoke breakage and or fire NOTE The battery pack includes an internal safety protec...

Страница 10: ...ed against vertically falling drops of water Antenna symbol for devices that include RF transmitters FCC ID FCC Identification code indicating traceability to FCC compliance Rx ONLY Reference to US FD...

Страница 11: ...ielectric breakdown ESD can be caused by a buildup of static electricity by tribocharging or by electrostatic induction At lower relative humidity as the environment is drier charge generation will in...

Страница 12: ...nt to the next to ensure the maximum level of hygiene and safety For a correct interpretation of a spirometry test the measured values must be compared either to the so called normal or predicted valu...

Страница 13: ...y Flow at 75 of FVC L s FEV05 Volume expired after 0 5 seconds L FEV05 FEV05 FVC x 100 FEV075 Volume expired after 0 75 seconds L FEV075 FEV075 FVC x 100 FEV2 Volume expired in the first 2 seconds of...

Страница 14: ...usable soft sensor for adults Reusable paediatric soft sensor Range SpO2 Arms Range SpO2 Arms Range SpO2 Arms 70 100 1 19 70 100 1 470 70 100 1 390 70 80 0 554 70 80 1 626 70 80 1 851 80 90 1 32 80 90...

Страница 15: ...octor 1 6 3 Oximetry alarms description Spirobank II is equipped with audio and visual alarm indicators to alert the operator to provide prompt patient attention or to abnormal device conditions Spiro...

Страница 16: ...h of the above mentioned oximetry sensors is used 1 6 4 Other features Memory Memory capacity for over 10000 spirometric tests The precise number depends on the individual configuration so it cannot b...

Страница 17: ...discharge protection 8kV contact 15kV air Magnetic field immunity 30 A m Radio Frequency Immunity 3V m 80 2700 MHz MIR will make available on request circuit diagrams component part lists description...

Страница 18: ...displayed with a reduction of the indicators 2 3 Main screen On the main screen the following areas can be accessed patient data management area spirometry area oximetry area archive area archive are...

Страница 19: ...e key and then the key The service menu shows the following list of items Change date time LCD settings Bluetooth settings Select language Delete memory Select standard Select predicted Select turbine...

Страница 20: ...standard to be used ATS ERS or NHANES III with the keys and then press the setting takes effect and the device returns to the Service Menu WARNING If the NHANES III standard is selected it is not pos...

Страница 21: ...hreshold or rise above the maximum threshold of the SpO2 and BPM values previously set Use keys and to decrease increase the values and select moving the arrow with the keys and At the end press to re...

Страница 22: ...g of the turbine is necessary The disposable turbine is check before the application of the external packaging for this reason it doesn t requires a periodic calibration If a calibration must be perfo...

Страница 23: ...robank II is calibrated correctly then the FVC syringe value will be 3 00 FVC x 1 026 BTPS 3 08 L FVC at BTPS If the ambient temperature is 20 C the FIVC syringe value will be 3 00 FIVC x 1 102 BTPS 3...

Страница 24: ...en without modifying any data WARNING A new patient is not created from the previous patient when selecting this function Patient info however can be modified Future tests will be associated to the pa...

Страница 25: ...efault settings of the PC software will be transferred to the device and will remain in the device even when used in stand alone mode for example when the Spirobank II is connected to the PC and the u...

Страница 26: ...results of the best test To end the test press During the test the Spirobank II emits beeps the frequency of which are directly proportional to the inspired and expired velocity of the air This helps...

Страница 27: ...rst screen displays a Flow Volume graph of the Forced Vital Capacity pressing it is be possible visualize the best value of FVC FEV1 FEV1 and PEF with the percentage change compared to the predicted v...

Страница 28: ...argest FVC is 100 mL 2 11 Oximetry Testing WARNING Check if the oximetry function is available in the device this function is an option in some models This is not available in BASIC version WARNING Th...

Страница 29: ...the Spirobank II timer resets itself and the device starts recording data For any type of oximetry test if the sensor is not properly connected the following message will be displayed on screen after...

Страница 30: ...ct to biocompatibility tests Choose an application site on the patient s finger or toe where the light source will be directly over and in line with the detector The preferred sites are the forefinger...

Страница 31: ...manage Spirobank II The user can perform on line wireless tests of spirometry or oximetry and the tests will be directly stored in the database of winspiroPRO software 3 2 PC connection via USB port W...

Страница 32: ...liquid solution and shake it so as to remove any impurities Leave the turbine immersed for the time specified in the instruction of the solution WARNING To avoid irreparable damage to the reusable tur...

Страница 33: ...an the sensor WARNING Do not sterilize by irradiation steam or by using ethylene oxide Unplug the sensor from the device before cleaning or disinfecting it The sensor included with the Spirobank II is...

Страница 34: ...t was not performed correctly Repeat the test and follow closely the indications shown on the screen Certain spirometry and or oximetry parameters are not shown at the end of a test Personalized param...

Страница 35: ...n If the product is utilised differently from the use described in the User s Manual If any alteration adjustment modification or repair has been carried out by personnel not authorised by the manufac...

Отзывы: