miniDSP Ltd, Hong Kong /
/ Features and specifications subject to change without prior notice
24
5.4
R
UNNING A MULTICHANNEL MEASUREMENT
Click on the Measure button. At the lower right of the measurement window, check that “Multiple inputs” is
selected and that your desired range of channels is selected.
If “Generate individual measurements” is
not
selected, as shown above, you will get a single measurement that
shows the range covered by all the measurements, and the average of all measurements:
The usual graph display controls work on this graph, such as limits and smoothing. (The example above uses
“Var smoothing.”)
If “Generate individual measurements”
is
selected, each microphone will produce a separate measurement. In
addition, the same measurement that shows the range covered by all the measurements and their average will
be produced. (In other words, up to 17 individual measurement graphs will be produced from one sweep.)