Steam Sterilizer

 

 

Operation Manual 

Please read the manual carefully, operate it according to the manual 
strictly, and avoid any hazard! 

 
 
 

 
 
 
 
 

 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
This manual applies to SUN18-

Ⅲ、

SUN23-

  (S)Steam Sterilizer.

 

 
File No

MTQM/C1-7.5.1-05-02/

-S(

)

10 

 
 
                                                                                                  REV1607ER 
 

 
 

Содержание SUN12-III

Страница 1: ...Operation Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN18 Ⅲ SUN23 Ⅲ S Steam Sterilizer File No MTQM C1 7 5 1 05 02 Ⅲ S E 10 REV1607ER ...

Страница 2: ...meter 1 4 Diagram Specification 2 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 5 8 Emergency Condition 9 9 Maintenance and Service 9 10 Transportation and Storage 11 11 Attention 12 Appendix NO 1 Circuit diagram of sterilizer NO 2 Steam loop ...

Страница 3: ... reliable product in sterilizing 1 Printer 2 USB 3 Used water exit 4 Water exit of water tank 5 Power 6 Power source socket 7 Fuse 2 Application Scope This machine can be applied in sterilizing for the tooth device operation device Sanitary Material and other heat resisting articles existing in the Dept of Stomatology Ophthalmology and lab When Choose 121 sterilization procedure the highest temper...

Страница 4: ... of sterilization etc 3 Tech Parameter Chamber size of the Steam Sterilizer Diameter X Depth 18L 249mmX355mm Chamber size of the Steam Sterilizer Diameter X Depth 23L 249mmX460mm Rated Voltage 230 50Hz Rated Power 1800VA 2300VA Sterilizing Pressure temp 100 120KPa 121ºC 200 220KPa 134ºC Both of above are Under Standard Atmospheric pressure Fuse Melting pipe 250V T 15A Capacity of water tank 4L Pro...

Страница 5: ... packing Dismantle thin film open the packing get the equipment from the packing Hauling precautions forbid lift up door forbid lift up legs of machine forbid Rollover or upside down Hauling sketch map is as follows 5 2 The air cooling window attached to the exterior surface of the Steam Sterilizer shall be non clog by dust or articles and the Steam Sterilizer shall working in the ...

Страница 6: ...ng working time keep it away from any combustible materials The equipment must be far away from flammable items 5 6 After placing the Equipment Check whether the power supply is in conformity with the requirements 230 5 0 hz voltage fluctuation range 10 power supply 2300 VA 23L 1800VA 18L In order to ensure the safety of the person and equipment must be equipped a ground wire the equipment casing ...

Страница 7: ... connect well to the power source Connect the power source hole under the back left side of the Steam Sterilizer and the power socket by the attached plug Press the green switch right bottom corner of the front side when the indicator light is on it means that Steam Sterilizer get through with the electric power and the process then is in initialization status and showing LD being on the screen At...

Страница 8: ... description 7 1 Display Interface 7 1 1 Starting Interface Press Enter to start 7 1 2 Menu Interface interchange the different interfaces with left buttons 7 1 3 Operation Interface Temp T Current temp inside of the Chamber ...

Страница 9: ...tion Code Icon Water Scarce Alarming with different error codes see the codes index 7 1 4 Setting Interface Water detect Enable the fuction or not Time and Date setting the printing time for the machine operations Setting Self definition program is only under special situation the unprofesional is not allowed also the sterilization effecitveness not guaranteed ...

Страница 10: ...leasing during the sterilization Please clean the filter in the same time with you draining the used water Note Be sure to fill distilled water 1 Before filling the distilled water you must turn on the switch of the power 2 Please fill the distilled water only after finish working 7 3 Function description Vacuum sterilizer divided into the following process preheating process the vacuum step up pr...

Страница 11: ...gain begin the second pre vacuum process 7 3 3 Sterilization process This trip is insulation packing process when the pot temperature dropped to the set sterilization temperature of 0 5 the pump starts pumping water to the steam generator the pot temperature rise to the set sterilization temperature of 0 9 open the exhaust exhaust valve allowing the pot temperature is kept between the set steriliz...

Страница 12: ...t possible to resolve please contact the dealers We will try our best to provide help to you in short time After the alarm The machine can t work Long press the start key can unlock alarm Must wait for the pressure display 0 Kpa can open the door there will be set aside heat and residual hot water Operator also need keep away from it prevent burns Error code and resolving measures NO Code display ...

Страница 13: ...u Fail on keeping pressure and temperature Inspect the steam leaking 14 EC long du The process interrupted artificially Long press the start key can unlock alarm 9 Maintenance and Service Shut off the power before service or components replacement and the service or replacement work shall be done by the technician Working environment minimum brightness to 215 15 lx 9 1 Clean the top water tank by ...

Страница 14: ...st Important the inset work must be done equably to the groove During the installation the laid four equal points of the ring must be inserted to the groove firstly later for the same work for the remaining segments of the ring After finishing press the sealing ring equally by the force of hand 4 Note The inner circle of the ring may be raised up when insert the ring into the groove Please press i...

Страница 15: ...ater collecting tank 4 is the water exiting mouth of the water storing tank Revolving the water exit button by anticlockwise direction so as to release the water 10 3 The autoclave shall be transferred and stored with the below conditions Ambiance temperature 40ºC 55ºC Relative humidity 85 Atmospheric pressure 500HPA 1060HPA Note Don t drag or drop during transportation 11 Attention ATTENTION 1 Th...

Страница 16: ...ms 16 The equipment must be reliable grounding 17 Must use the attached plug do not change it 18 When Choose 121 sterilization procedure the highest temperature in cavity body is 125 When choose 134 sterilization procedure the highest temperature in cavity body is 138 19 This product is not suitable for sealing not resistant to high temperature products sterilization 20 This product is not suitabl...

Страница 17: ...15 APPENDIX NO 1 Circuit diagram ...

Страница 18: ...16 NO 2 Steam Loop ...

Страница 19: ... Co Ltd Yunzhen Road Industrial Zone Yunlong Town Yinzhou Ningbo City China Tel 86 0574 88320096 Fax 86 0574 88320098 Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Tel 49 40 2513175 Fax 49 40 255726 ...