9-5. Writing Calibration Data
Calibration ZERO, +SPAN, and -SPAN AD values can be adjusted.
(1) Click [Calibration]. The button changes to [Editing].
Calibration items can be set in this Editing mode.
(2) Adjust the values of the calibration items to set.
(3) Click [Start] after changing calibration items.
(4) The following message appears. Click [Yes].
Calibration item writing is complete.