10
Output fluctuates
abnormally.
No
1
Inform Minebea about the contents
of failure and situation in details
Yes
No
The connected tie-bar gage was
broken or the signal path (signal
conductor) might be
disconnected.
Approx. 5V is output.
No
(1) Please confirm the condition of the connector
for I/O.
(2) Please confirm the operation of voltage output
by check function. (Refer to paragraph 6.)
Please remove the cable of tie-bar gage.
Measure the resistance value between 1 pin and 4 pin
of the connector for tie-bar gage side.
(Please set the connection of the measuring
instrument to ohm range.)
Measure the resistance value between 7 pin and 2 pin
of the connector for tie-bar gage side.
Measure the resistance value between 4 pin and 5 pin
of the connector for tie-bar gage side.
Measure the resistance value between 2 pin and 3 pin
of the connector for tie-bar gage side.
Voltage is output.
No
Yes
・
Between 1 pin - 4 pin=350
Ω
・
Between 7 pin - 2 pin=350
Ω
・
Between 4 pin - 5 pin=0
Ω
・
Between 2 pin - 3 pin=0
Ω
Yes
Inform Minebea about the contents of
failure and situation in details
Yes
Содержание CSA-593