33
6-7. Entering load cell output voltage values to register zero point
Write the data in the following table to enter a numerical value for registering the zero point. Values can be
registered in 0.00001mV/v increments.
Command no.
(RWwnE): 4 stations shared
(RWwn6): 2 stations shared
General data area
(RWwnC~RWwnD): 4 stations shared
(RWwn4~RWwn5): 2 stations shared
9008
Data range: -250000~250000
(-2.5 mV/V~2.5 mV/V)
TE-L error
The following codes are set if the load cell output value set (in mV/V) is less than -2.5 mV/V and exceeds
the zero adjustment range on the negative side:
Status
Error code
Auxiliary error code
TE-L error
4
0
TE-H error
The following codes are set if the load cell output value set (in mV/V) exceeds 2.5 mV/V and the zero
adjustment range on the positive side:
Status
Error code
Auxiliary error code
TE-H error
4
1
Unexecutable error
The following codes are set if performed in modes other than calibration mode or if not set according to the
flowchart at the beginning of this chapter:
Status
Error code
Auxiliary error code
Unexecutable error
99
2