5.2.1.4
Negative weighing value
Negative stable netto weight value is displayed like this:
5.2.2
Counting Mode
Counting items based on the weight of a reference sample count.
Example: Determine the number of parts; weigh in the selectable reference sample
quantity (in this example: 20).
Release the button. The application is activated.
The stored reference quantity from the initialization before is displayed. The number
of selectable samples is blinking.
1.
To select application Count, press the
Mode
F
button, until [COUNT] is displayed with
application icon
at bottom of the screen.
2.
If there is already a piece weight initialized, [CLR.PW] is displayed. Otherwise [PUT
20] (for example) is displayed, then go on at step 4.
3.
Press the
Unit
(No) button to use the stored reference weight.
4. Press the
Off
(Yes) button to clear the stored PW (average piece weight) and set a
new one.
5.
[PUT 20] (for example) is displayed.
6. Choose the desired reference quantity sample by pressing the
Mode
F
button briefly
(decrement) or the
Unit
(No) button briefly (increment) to toggle the choices (10, 20,
50, 100, 200).
7.
Press the
Off
(Yes) button to confirm reference sample quantity.
8. If the pan is empty, then [PUT.PW] (put reference) is shown on the display. Put the
desired reference quantity on the pan and go on at step 7.