If there are no reference samples put on the pan or the weight is smaller than 2d or
the calculated piece weight would be smaller than 2d/10, the display will show
[REF.ERR] for about 2 seconds and then goes back to step 9.
If the weight on the pan is < reference quantity * d AND also calculated piece weight ≥
(2d/10), then the display will show[LOW.REF] for about 2 seconds.
Example: PUT 50 ; d = 0.2 g ; weight on the pan = 2.0 g ; then [LOW.REF] is displayed.
But is the weight on the pan = 1.8 g, then [REF.ERR] is displayed.
9. Press the
Off
(Yes) button to store the reference quantity sample.
The display shows [PW.OK] (piece weight ok) for 2 seconds. Go on step 10.
10. Counting is initialized. On the display, the actual number of pieces is shown: [20
pcs]for example.
11. Replace the reference sample with the test sample.
The display shows the quantity of the sample count.
12. Press the
Mode
F
button briefly.
5 Operation
Floor scale Puro®
Minebea Intec
EN-15