Connect the screens to the grounding terminal or the screen clamping rail of the
device; see Chapter
.
-
Establish grounding/equipotential bonding between devices/system components;
see Chapter
4.3 EMC-compliant installation
4.3.1
Connecting the screens to the screen clamping rail
The screens must be connected to the screen clamping rail using cable clamps as pictured
or the supplied screen terminals.
4.3.2
Connecting the equipotential bonding conductor
1
The equipotential bonding conductor (1) must be connected as pictured.
Process Controller Maxxis 4 PR 5500
4 Device installation
EN-22
Minebea Intec