— 7 EVEN
7 data bits, even parity
— 7 Odd
7 data bits, odd parity
— 7 NONE
7 data bits, no parity
— 8 NONE
8 date bits, no parity*
— STOP
Stop Bit
— 1
1*
— 2
2
— RESET
Factory setting
— NO
not restored*
— YES
enabled
— END
Leave menu level
— MODE
PC Output Mode
— OFF
disabled*
— MAN.OUT
Manual output without stability = device
expects a request via a serial interface.
Response is generated
immediately
— MAN.STA
Manual output without stability= device
expects a request via a serial interface.
— INT.OUT
interval output with 16 characters
without
stability = device sends weighing values
automatically without stability in the selected
interval of display refresh cycles (see below)
— AUT.OUT
automatic output
without
stability with 16
characters = device sends weighing values
without stability automatically
every
display
refresh cycle
— AUT.STA
automatic output with stability
with
16
characters at each value change = device sends
weighing values with stability automatically
— INTERV
Set the output interval (when INT.OUT is
selected)
— 1 CYC
Each display cycle
— 2 CYC
After 2 display cycle
— 5 CYC
After 5 display cycle
— 10 CYC
After 10 display cycle*
— 20 CYC
After 20 display cycle
— 50 CYC
After 50 display cycle
— 100 CYC
After 100 display cycle
— RESET
Factory setting
— NO
not restored*
— YES
enabled
— END
Leave menu level
7.4.6
Menu Selection [PC.OUT]
Enter this menu to define PC Output parameters.
This is an additional interface based on RS232 with USB-C connector.
Default settings are identified by an "*"
PC.OUT
"Without stability" means here weight values with or without stability.
"With stability" means here only weigh values with stability!
Compact Scale Puro® Count
7 Menu Settings
EN-52
Minebea Intec