Note:
-
Zero compensation may not be accepted when the setting of power-on-zero clear is
[EFFECTIVE], and the condition at power-on is as follows;
-
When the weighing is not in the stable condition, the display becomes [-888
888].
-
When the weighing value exceeds over ±10 % of the weighing capacity, the
display becomes [- --- ---].
-
When the power-on-zero is applied, each data of tare weight cancellation, preset
tare weight cancellation, zero set and zero tracking is cleared.
9.13 Power-on tare weight cancellation
When the power turns on, power-on-tare-weight-cancellation processes the tare weight
cancellation.
When the power turns on, the power-on-tare-weight-cancellation-clear clears the tare
weight and returns the display to the gross weight.
9.13.1
Operation of power-on-tare-weight-cancellation (TARE AT POWER ON)
The operation of power-on-tare-weight-cancellation is selectable from [EFFECTIVE] or
[INVALID].
[INVALID] is selected as default.
9.13.2
Power-on-tare-weight-cancellation-clear (TARE CLR. AT POWER ON)
The operation of power-on-tare-weight-cancellation-clear is selectable from [EFFECTIVE]
or [INVALID].
[INVALID] is selected as default.
9.14 Operation of hold (OPERATION OF HOLD)
The operational condition of hold is selectable from [SYNC.WITH HOLD] or [SYNC.WITH
BATCH FINISH].
[SYNC.WITH HOLD] is selected as default.
Note:
Target of hold operation is applied in Chapter
and
.
9.15 Print command (PRINT COMMAND)
The print command is selectable from [IN STABLE MODE] or [ANYTIME] when the print
command is input from the external I/O.
[IN STABLE MODE] is selected as default.
When [ANYTIME] is selected, the print command is always accepted.
And when [IN STABLE MODE] is selected, the print command is accepted only when
[STABLE] sign of the status display-1 lights.
Weight Controller CSD-918
9 Basic function setting display (BASIC FUNCTION)
EN-82
Minebea Intec