The default setting is [DISP.ACC] (3.9.1) "display resolution." The resolution is increased
when [10 FOLD] (3.9.2) "10-fold" or [100FOLD] (3.9.3) "100-fold" is selected. "10-fold"
increases the resolution of the net value by one step (display resolution x 10), "100-fold"
increases it two steps (display resolution x 100).
Setting: [APPLIC.] - [APPLIC.1] - [NEUTR.M] - [RESOLUT] (3.9) (siehe auch Kapitel
or
).
7.8.5.3
Decimal Places
In neutral measurement, not only whole numbers but also decimal numbers (for example,
1.25 o electrical cabling) can be displayed [DEC.PLCS] (3.10) "Decimal Places". The number
of decimal places displayed can range from none up to 3 places.
Setting: [APPLIC.] - [APPLIC.1] - [NEUTR.M] - [DEC.PLCS] (3.10) (see chapter
or
).
7.8.5.4
Saving Weight Values
The weight on the platform is saved as a reference value as soon as the platform has
stabilized [SAVE WT] (3.11) "Parameter for saving weight values". "Stability" is deined as
the point at which the luctuation of a measured value lies within a deined tolerance. The
narrower the tolerance, the more stable the platform is at "stability".
The [ACC.STAB] (3.11.2) "increased stability" setting has a lower tolerance so that the
average piece weight saves is more accurate and the results more reproducible; however,
the measurement time can take longer.
Setting: [APPLIC.] - [APPLIC.1] - [NEUTR.M] - [SAVE WT] (3.11) (see chapter
or
7.8.5.5
Measuring with Two Weighing Platforms
You can use two weighing platforms simultaneously with the Neutral Measurement
application. When using two platforms, you can choose from the following operating
modes:
-
Counting with two platforms of the same type (see chapter
).
-
Counting with one reference platform and one weighing platform (see
chapter
7.8.5.5.1 Neutral measurement with two weighing platforms
Use this mode to count diferent types of sample material with diferent weights. For
example, count the lighter-weight pieces on one platform and the heavier pieces on
another. You can deine one of the two scales as the reference scale.
The reference scale is the irst scale active when you switch on the device, regardless of
the setting for automatic initialization of the application.
7.8.5.5.2 Neutral measurement with one reference platform and one weighing platform
In this operating mode, the reference platform is a high-resolution weighing platform
with a relatively low maximum capacity. The other platform is used for weighing heavier
samples, and has a high capacity with a relatively low resolution. This allows you to both
determine the reference sample weight with high resolution; i.e., very precisely, and to
count large amounts of parts, without requiring an expensive high-resolution, high-
capacity weighing platform.
The system can be conigured to switch automatically to the reference platform for
initialization (the measured value line shows [REF]). Following initialization, you can
switch to the counting platform.
Setting: [APPLIC.] - [APPLIC.1] - [NEUTR.M] - [REF.WP] (3.13) (see chapter
or
7 Operating
Combics 2 Series Complete Scales Models CAW2P, CAW2S
Minebea Intec
EN-123
Содержание CAW2P
Страница 205: ...Typ Waage Minecomb Typ Auswerteger t TA EG Bauartzulassung T11379 Pr fschein D09 11 02 Typ Minecomb...
Страница 206: ...Typ Waage Minecomb Typ Auswerteger t TA EG Bauartzulassung T11379 Pr fschein D09 11 02...
Страница 207: ...Option Y2 Typ Waage Minecomb Typ Auswerteger t TA EG Bauartzulassung T11379 Pr fschein D09 11 02...
Страница 209: ...CAW3S1 60FE NCE Typ Waage Minecomb Typ Auswerteger t TA EG Bauartzulassung T11379 Pr fschein D09 11 02...
Страница 210: ......
Страница 211: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 220: ......
Страница 222: ......
Страница 223: ......