7.8 Neutral Measurement nM application
With this "Neutral Measurement"
application you can measure the length, surface and
volume of parts that have roughly the same speciic weight. The "o" symbol is displayed
as the unit.
Menu: [APPLIC.] - [APPLIC.1] - [NEUTR.M] (see chapter
).
Features
-
Save the reference weight "wRef" from the weighing platform
-
Enter the reference weight "wRef" through the keypad
-
Enter the factor for calculation "nRef" using the keypad
-
Enter reference sample weight using a barcode scanner
-
Measuring with two weighing platforms
-
Activate Info mode via the
key
-
Toggle the display between measurement and weight via the
key
-
The level of accuracy (display resolution) can be set when the calculated reference
weight is applied
-
Automatic taring of container weight
Setting: [APPLIC.] - [AUT.TARE] (3.7) (see chapter
-
Automatic initialization when the scale is switched on. The indicator is initialized with
the most recently used values for reference sample quantity "nRef" and reference
sample weight "wRef."
Setting: [APPLIC.] - [AUT.STRT] (3.8) (see chapter
).
7.8.1
Exit application, delete parameters
The value of the reference sample weight in the reference memory remains active until it
is deleted via the
key or overwritten or until the application is changed.
The reference sample weight also remains saved after the scale is turned of. You can
assign diferent functions to the
key for deleting applications. When you clear
applications, you can delete either the data stored for all applications or just selected data
stored for the active application.
Setting: [APPLIC.] - [CLER.CF] (3.24) (see chapter
7.8.2
Tare function
If you store a tare (weight value) by pressing the
key, you can later enter a tare value
manually. The tare value you enter is added to the stored tare value.
Setting: [APPLIC.] - [TARE.FNC] - [NORMAL] (3.25.1) (factory setting) (see chapter
).
A tare value entered manually overwrites a stored tare value (weight value). If you enter a
tare value manually, a tare value (weight value) stored later overwrites the manually
entered value.
Setting: [APPLIC.] - [TARE.FNC] - [SPECIAL] (3.25.2) (see chapter
7.8.3
Restore factory default settings
Setting: [APPLIC.] - [RESET] (9.1) (see chapter
).
Combics 2 Series Complete Scales Models CAW2P, CAW2S
7 Operating
EN-120
Minebea Intec
Содержание CAW2P
Страница 205: ...Typ Waage Minecomb Typ Auswerteger t TA EG Bauartzulassung T11379 Pr fschein D09 11 02 Typ Minecomb...
Страница 206: ...Typ Waage Minecomb Typ Auswerteger t TA EG Bauartzulassung T11379 Pr fschein D09 11 02...
Страница 207: ...Option Y2 Typ Waage Minecomb Typ Auswerteger t TA EG Bauartzulassung T11379 Pr fschein D09 11 02...
Страница 209: ...CAW3S1 60FE NCE Typ Waage Minecomb Typ Auswerteger t TA EG Bauartzulassung T11379 Pr fschein D09 11 02...
Страница 210: ......
Страница 211: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 220: ......
Страница 222: ......
Страница 223: ......