Product Specifications
3-8
Range and accuracy for
PR
measurement:
Range Accuracy
20 - 250 BPM
±3 BPM
251 - 300 BPM
Undefined
Updating cycle of SpO
2
values:
1s
3.2.10 TEMP Specification
Parameter Specification
Displayed parameter
TEMP
Measurement range
25 - 44 ºC
(
77 - 111.2 ºF
)
Resolution
In MONITOR mode: 0.1ºC (0.2 ºF)
Accuracy
In MONITOR mode: 25 - 32 ºC (77 - 89.6 ºF):
±
0.2 ºC (
±
0.3
ºF) including 32 ºC
(
89.6 ºF
)
In MONITOR mode: 32 - 44 ºC (89.6 - 111.2 ºF):
±
0.1ºC
(±
0.2 ºF
)
excluding 32
ºC
(
89.6 ºF
)
Typical prediction time
< 12 s