Quick Reference Guide
V
V
V
P/N: 0002-08-50038-00 Rev A
25
To adjust a patient’s arrhythmia threshold setting on the V12 or V21 bedside
monitor:
a. Select ECG waveform to enter
ECG Setup
tab
b. Select
Arrhy Setup
tab
c. Select individual arrhythmia threshold
d. Scroll and select value or enter value using on screen keypad
e. Select
Accept
to save settings and return to the main screen
Or
a. Select
Alarms
tab from Navigation Area
b. Select
Arrhythmia
to enter the
Arrhythmia
tab
c. Select
Setup
d. Select individual arrhythmia threshold
e. Scroll and select value or enter value using on screen keypad
f. Select
Accept
to save settings and return to the main screen
RESET ALARMS
(Version 2.5 or greater)
Select
Alarm Acknowledge
in the Navigation Area on the V12 or V21 bedside
monitor to silence all current alarms for a preconfigured time interval or until
the alarm condition is resolved.
Panorama® Central Station interface:
Selecting
Alarm Acknowledge
on the
V12 or V21 bedside monitor will silence an alarm on the Panorama Central
Station, WorkStation, and ViewStation for a
preconfigured amount of time
.
Selecting
Mute
at the Panorama Central Station
will not silence
an alarm at the
V12 or V21 bedside monitor.
PAUSE ALL ALARMS
(Version 2.5 or greater)
Select
Audio Pause
in the Navigation area on the V12 or V21 bedside monitor
to silence all current and future alarms for a preconfigured time interval.
Panorama® Central Station interface:
Selecting
Audio Pause
on the V12
or V21 bedside monitor will pause active alarms on the Panorama Central
Station, WorkStation, and ViewStation for a
preconfigured amount of time
.
Selecting
Mute
at the Panorama Central Station
will not
pause alarms on the
V12 or V21 bedside monitor.