SpO
2
Measurement
Specifications
2 - 6
2.4
SpO
2
Measurement
When configured with the DPM SpO
2
module, NELLCOR NELL-3 SpO
2
module, or MASIMO
MS-2013 SpO
2
module, the Accutorr V can perform SpO
2
measurements.
The SpO
2
measurement function complies with the requirements of ISO9919.
2.4.1
DPM SpO
2
Module Performance
The DPM SpO
2
accuracy has been validated in human studies against arterial blood sample
reference measured with a CO-oximeter. Pulse oximeter measurements are statistically
distributed, and only about two-thirds of the measurements can be expected to fall within the
specified accuracy compared to CO-oximeter measurements.
2.4.1.1
SpO
2
Range:
........................................................................................... 0
–
100%
Resolution:
.................................................................................... 1%
Accuracy in the static condition:
2.4.1.2
Pulse Rate
Range:
........................................................................................... 20
–
254 BPM
Resolution:
.................................................................................... 1 bpm
Accuracy in the static condition:
2.4.1.3
Alarm setting range
PATIENT TYPE
SPO
2
RANGE
70% TO 100%
0 – 69%
Adult or pediatric
±2 digits
Undefined
Neonate
±3 digits
Undefined
PATIENT TYPE
PR RANGE
ACCURACY
Adult / Pediatric / Neonate
20 – 254 bpm
± 3 digits
SPO
2
MODULE
SPO
2
PR
DPM SpO
2
module
50 – 100%
0 – 254 bpm