NOTE:
•
Guidance and precautions for even movement:
–
Make sure there is enough coupling gel along the scan path.
–
Always move the probe slowly and steadily.
–
Continuous contact is required throughout the length of the extended image. Do not lift
the probe from the skin's surface.
–
Always keep the probe perpendicular to the skin's surface. Do not rock, rotate or tilt the
probe during the scan.
–
The system accommodates a reasonable range of motion velocity. Do not make abrupt
changes in motion speed.
–
Deeper scans generally require reduced acquisition speed.
•
Needle mark cannot be displayed in iScape imaging mode.
6.11.1 Basic Procedures
Perform the following procedure:
1.
Connect an appropriate iScape compatible probe. Make sure there is enough coupling gel
along the scan path.
2.
Press the user-defined key for <iScape View> or tap [iScape] on the touch screen (it is
available after enter Power/Color mode).
3.
Optimize the 2D mode image:
In the capture preparation status, tap [B] ([Power]/[Color]) page tab to go for the B mode
image optimization. Do measurement or add comment/bodymark to the image if needed.
4.
Tap [Start Capture] or press <Update> on the control panel to begin the capture.
5.
Scan slowly to obtain a single extended field of view image. You can also erase and retrace if
the image is not satisfactory.
–
During image acquisition, none of the parameters are adjustable, and functions such as
measurement, comments and body marks are not available.
–
A green box on the image indicating the boundary between the merged images and the
unfinished images.
–
During image slicing, the system gives feedback on the probe’s moving speed in the form
of colors and words. The meanings are as follows:
6.
The system enters into image review status when the acquisition is completed. You can
perform operations such as parameter adjusting.
7.
Do one of following to end image capture:
–
Tap [Stop Capture] on the touch screen.
–
Press <Update>.
–
Wait until the acquisition completes automatically.
Status
ROI Color
Tip
Speed too low
Blue
Moving speed of the probe is too low.
Appropriate
Green
/
Speed too high
Red
Moving speed of the probe is too high.
Содержание Imagyn 7
Страница 2: ......
Страница 14: ...This page intentionally left blank...
Страница 20: ...This page intentionally left blank...
Страница 54: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...
Страница 118: ...This page intentionally left blank...
Страница 126: ...This page intentionally left blank...
Страница 196: ...This page intentionally left blank...
Страница 240: ...This page intentionally left blank...
Страница 280: ...This page intentionally left blank...
Страница 298: ...This page intentionally left blank...
Страница 406: ...This page intentionally left blank...
Страница 416: ...This page intentionally left blank...
Страница 491: ......
Страница 492: ...P N 046 019593 01 3 0...