D - 2
Note: When arrhythmia alarms occur, check the patient’s condition and the ECG connections.
D.1.3
ST Physiological Alarm Messages
Note: XX represents the ECG lead label.
D.1.4
Resp Physiological Alarm Messages
D.1.5
SpO
2
Physiological Alarm Messages
D.1.6
PR Physiological Alarm Messages
Missed Beat
Prompt
A-Fib
Prompt
Alarm message
Default priority
ST alarm mode
Alarm messages
Default priority
Cause and solution
Absolute
ST-XX High
Med
The ST value of respective ECG lead has risen above the
high alarm limit or fallen below the low alarm limit. Check
the patient’s condition and check if the patient category
and alarm limit settings are correct.
ST-XX Low
Med
Relative
ST Single
Med
ST value of any ECG leads has risen above the high alarm
limit or fallen below the low alarm limit. Check the patient’s
condition and check if the patient category and alarm limit
settings are correct.
ST Dual
Med
ST values of two or more ECG leads have risen above the
high alarm limit or fallen below the low alarm limit. Check
the patient’s condition and check if the patient category
and alarm limit settings are correct.
Alarm message
Default priority
Cause and solution
Resp Artifact
High
The patient’s heartbeat has interfered with his respiration. Check the
patient’s condition and the Resp connections.
Apnea
High
The respiration signal was so weak that the monitor cannot perform
respiration analysis. Check the patient’s condition, module and patient
connections.
Alarm message
Default priority
Cause and solution
SpO2 Desat
High
The SpO
2
value falls below the desaturation alarm limit. Check the
patient’s condition and check if the alarm limit settings are correct.
Alarm message
Default priority
Cause and solution
No Pulse
High
The pulse signal was so weak that the monitor cannot perform pulse
analysis. Check the patient’s condition, SpO2 sensor and measurement
site.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...