22 - 5
22.1.5
Changing EWS Settings
22.1.5.1
Changing the Scoring Protocol
The monitor is configured with a default scoring protocol. To change the scoring protocol, follow this procedure:
1.
From the EWS page select
Setup
.
2.
Set
Score
.
22.1.5.2
Setting the Scoring Confirmation Switch
To select if confirmation is required before saving score, follow this procedure:
1.
From the EWS page select
Setup
.
2.
Set
Scoring Confirmation
switch.
◆
Off: the monitor automatically saves the scoring result after the scoring is completed.
◆
On: you need to confirm that whether the scoring result is saved or not after the scoring is completed.
22.1.5.3
Setting the Manual Data Timeout
The manually input parameter data becomes invalid after a preset time. To set the timeout period for the input
data, follow this procedure:
1.
From the EWS screen select
Setup
.
2.
From the
Manual Data Timeout
area, select a desired parameter and set its timeout period.
NOTE
•
If the data is expired and not updated, the monitor displays the corresponding parameter score in
outline font, and gives a timeout alarm.
22.1.5.4
Setting Auto Scoring
The monitor automatically starts scoring at the preset interval. To set auto scoring, follow this procedure:
1.
From the EWS page select
Setup
.
2.
Set
Auto Scoring
:
◆
Interval
: the monitor automatically starts scoring at the preset interval.
◆
NIBP
: the monitor automatically starts scoring at the completion of each NIBP measurement.
◆
Alarm
: the monitor automatically starts scoring when an alarm occurs to the parameter for scoring.
◆
If no option is selected, the monitor does not initiate auto scoring.
22.1.5.5
Setting Auto Scoring Interval
1.
From the EWS page select
Setup
.
2.
Set
Interval
:
◆
By Score
: the monitor automatically starts scoring as per the interval selected for corresponding total
score.
◆
5 min
-
24 h
: If
Auto Scoring
is set to
Interval
, the monitor automatically starts scoring as per the
selected interval. If
Auto Scoring
is not set to
Interval
, the countdown timer of manual scoring is
selected.
22.1.6
Viewing Historical Scores
From the EWS screen, you can view the total score or subscores of the recent 24 hours. To do so, choose either of
the following ways:
■
Select the history total score area.
■
From the history total score area, swipe up with a finger.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...