B-3
Surge IEC
61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not apply
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
<5 % U
T
(>95 % dip in
U
T
) for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in
U
T
) for 5 s
Not apply
/
Power frequency
(50/60 HZ)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Guidance and Declaration - Electromagnetic Immunity
The device is intended for use in the specified electromagnetic environment. The customer or the user of the
device should assure that it is used in such an environment as described below.
Immunity test IEC60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conduced RF
IEC61000-4-6
3 Vrms
150 kHz to 80
MHz
3 Vrms
Portable and mobile RF communications
equipment should be used no closer to any part
of the system, including cables, than the
recommended separation distance calculated
from the equation appropriate for the frequency
of the transmitter. Recommended separation
distances:
P
d
2
.
1
Radiated RF
IEC61000-4-3
3V/m
80MHz to
2.5GHz
3V/m Recommended
separation
distances:
80 MHz
~
800 MHz
P
d
2
.
1
800MHz-2.5GHz