CAUTION:
11. The cover contains natural rubber latex and talc that can cause allergic
reactions in some individuals.
12. Please use the ultrasound gel compliant with the relevant local
regulations.
13. Do not connect this system to outlets with the same circuit breakers and
fuses that control the current of devices such as life-support systems. If
this system malfunctions and generates overcurrent, or when there is an
instantaneous current at power ON, the circuit breakers and fuses of the
building’s supply circuit may be tripped.
NOTE:
1. Do not use the system in the vicinity of strong electromagnetic field (such as a
transformer), which may affect the performance of the system.
2. Do not use the system in the vicinity of high-frequency radiation source, which may
affect the performance of the system or even lead to failure.
3. When using or placing the system, keep the system horizontal to avoid disbalance.
4. To avoid damaging the system, do not use it in following environment:
(1) Locations exposed to direct sunlight;
(2) Locations subject to sudden changes in environmental temperature;
(3) Dusty locations;
(4) Locations subject to vibration;
(5) Locations near heat generators;
(6) Locations with high humidity.
5. Turn ON the system only after the power has been turned OFF for more than 20
seconds. If the system is turned ON immediately after being turned OFF, the system
may not reboot properly and could malfunction.
6. Do not disconnect a transducer that remains in a live imaging state. This can
damage the system and / or transducer. Press [Freeze] key to freeze an image or
turn off the power of the system before connecting or disconnecting a transducer.
7. Remove ultrasound gel from the face of a transducer when the examination is
complete. Water in the gel may enter the acoustic lens and adversely affect or
damage the transducer array and lens.
8. You should properly back up the system to a secure external storage media,
including system configuration and settings and patient data. Data stored to the
system’s hard drive may be lost due to system failure, improper operation or
accident.
9. Do not apply external force to the control panel. The system may be damaged.
10. If the system is used in a small room, the room temperature may rise. Provide
proper ventilation and free air exchange.
11. To dispose of the system or any part, contact Mindray Customer Service Department
or sales representative. Mindray is not responsible for any system content or
accessories that have been discarded improperly.
12. To ensure optimal system operations, it is recommended that you maintain the
system under a Mindray service agreement. Various aspects of system performance
and operation can be maintained under the professional supervision of a service
representative.
13. Do not use gel, disinfectant, transducers or needle-guided brackets that are not
compatible with the system.
Safety Precautions IX
Содержание DP-8300
Страница 1: ...DP 8500 DP 8300 Digital Ultrasonic Diagnostic Imaging System Operator s Manual Basic Volume ...
Страница 2: ......
Страница 8: ......
Страница 20: ......
Страница 36: ......
Страница 54: ......
Страница 58: ......
Страница 82: ......
Страница 86: ......
Страница 92: ......
Страница 94: ......
Страница 96: ......
Страница 101: ...P N 046 000308 00 V1 1 ...