Functional Tests
Periodic Maintenance
6 - 30
6.8.7.8
PCV Adult Ventilation Mode Test
1.
Attach a breathing circuit and breathing bag.
NOTE:
For testing purposes always use a reusable breathing
circuit.
2.
Attach an adult test lung to the Y-fitting of the breathing circuit.
3.
Attach a Vent Tester between the EXP port and the expiratory hose.
4.
Set the O
2
flow to 3 L/min and set the N
2
O and AIR flow rates to minimum flow.
5.
Set the ventilator controls to:
6.
Select
PCV
to begin ventilation.
7.
Verify the Peak Pressure reading of the display is within ±4 cmH
2
O of the Peak Pressure
measured with the Vent Tester.
8.
Verify that the pressure waveform, Tidal Volume, Resp. Rate and minute volume values
appear on the screen.
9.
Verify that the PEAK Value reaches 20 ±4 cmH
2
O within five breaths from the start of
ventilation.
VENTILATOR CONTROLS
VENTILATOR SETTINGS
Patient Type
Adult
Ventilation Mode
PCV
Target Pressure -
P
TARGET
20
Breath Rate -
freq
8
I:E Ratio -
I:E
1:2
PEEP -
PEEP
Off
Inspiratory Slope -
T
slope
0.5
Содержание Datascope AS 3000
Страница 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Страница 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 127: ...Calibration Procedures Calibration 5 14 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Страница 178: ...AS3000 Service Manual 0070 10 0683 7 3 Warranty Manufacturer s Responsibility This page intentionally left blank...
Страница 179: ...0070 10 0683 Rev K August 8 2014...