6 - 5
Periodic Maintenance
Check Valve Cleaning
6.6.2
Cleaning Procedure
1. Remove Absorber Locking Mechanism and Absorber Mount Assembly:
1.
Rotate the cam handle to the unlocked position.
2.
Remove (2) Absorber Canister Assemblies.
3.
Remove Absorber Hose from port on Absorber Mount Assembly.
4.
Remove (2) 4mm hex screws securing the Absorber Locking Mechanism to the rods.
5.
Remove the Absorber Locking Mechanism and Absorber Mount Assembly.
FIGURE 6-2
Absorber Assembly
Absorber Mount
Assembly
Absorber Canister
Assembly
Absorber
Locking
Mechanism
Absorber
Hose