Monitoring NIBP
Configuring the NIBP Setup
BP10 Operator’s Manual
6 - 5
6.4.2
Taking a Measurement
1.
Prepare and make sure the patient is in a correct position. Refer to
for details.
2.
Turn on BP10.
3.
Verify that the patient category is correct.
4.
If not, press the
key, and then select
Patient
Info
→
Patient Category
and
set the patient size to
Adult
or
Pediatric
.Connect the extension air hose to the
NIBP connector on BP10.
5.
Select the appropriate sized cuff by referring to the limb circumference marked on
the cuff.
The width of the cuff should be 40% of the limb circumference, or 2/3 of the upper
arm’s length. The inflatable part of the cuff should be long enough to encircle at
least 50% to 80% of the limb.
6.
Apply the cuff to the patient’s upper arm at the same level as the heart, and make
sure the Φ marking on the cuff is aligned with the artery.
Do not wrap the cuff too tightly around the limb. It may cause discoloration, and
ischemia of the extremities.
Make sure that the edge of the cuff is within the marked range. If it is not, use a cuff
that fits properly.
7.
Connect the cuff to the extension air hose.
Avoid compression or restriction of NIBP hose. Air must pass unrestricted through
the tubing.
6.5
Configuring the NIBP Setup
The NIBP setup includes two levels settings. One group of settings are passcode
protected, the other group of settings are not.
Содержание BP10 NIBP Module
Страница 1: ...BP10 NIBP Module Operator s Manual...
Страница 2: ......
Страница 11: ...BP10 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 5...
Страница 16: ...Equipment Symbols Safety 1 6 BP10 Operator s Manual This page intentionally left blank...
Страница 38: ...Service Menu Configuring BP10 4 6 BP10 Operator s Manual This page intentionally left blank...
Страница 42: ...Discharging the Patient Patient Management 5 4 BP10 Operator s Manual This page intentionally left blank...
Страница 62: ...Disposing of the Batteries Battery 8 8 BP10 Operator s Manual This page intentionally left blank...
Страница 73: ...BP10 Operator s Manual 11 1 11 Accessories NIBP Accessories 11 2 Miscellaneous 11 3...
Страница 76: ...Miscellaneous Accessories 11 4 BP10 Operator s Manual This page intentionally left blank...
Страница 87: ...BP10 Operator s Manual B 1 B EMC...
Страница 93: ...BP10 Operator s Manual C 1 C FCC Compliance...
Страница 95: ...BP10 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3...
Страница 98: ...Abbreviations Symbols and Abbreviations D 4 BP10 Operator s Manual P power PR pulse rate...
Страница 99: ......
Страница 100: ...P N 046 011008 00 4 0...