Cleaning the Thermal Print Head
1.
Turn off the device.
2.
Open the recorder door with the latch.
3.
Remove the recording paper.
4.
Gently wipe the print head with cotton swabs, dampened with ethanol, to remove the dust and
foreign particles.
5.
Wipe off excess moisture with dry cotton swabs.
6.
When the print head is completely air dried, reload the recording paper and close the recorder
door.
Printing quality may be affected by a dirty print head. Clean the print head once per month, or as
needed. Check the printout to ensure the printing is legible and dark. Light printing may indicate a
dirty print head.
Disinfecting
Disinfection may cause damage to the equipment and is therefore not recommended for this device,
unless otherwise indicated by your hospital’s servicing schedule. Cleaning the device before
disinfecting is recommended.
The recommended disinfectants for the equipment include:
Ethanol 75%
Isopropanol 70%
Perform®
classic concentrate OXY
For recommended disinfectant agents for the reusable accessories refer to the instructions for use,
delivered with the accessories.
Sterilisation
Unless otherwise specified in the instructions for specific accessories, do not sterilise the equipment.