B - 2
Defibrillator/Monitor Operator’s Manual
B.2
Mindray Shockable Rhythm Analysis Algorithm Performance
Test results on the performance of the equipment configured with Mindray shockable rhythm analysis algorithm
meet IEC 60601-2-4 requirements
b
and AHA recommendations
a
.
Test results on IEC 60601-2-4 requirements are shown below.
Test results on AHA recommendations are shown below.
a
. Kerber RE, et al, “Automatic External Defibrillators for Public Access Defibrillation: Recommendations for
Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms, and
Enhancing Safety: A Statement for Health Professionals from the American Heart Association Task Force on
Automatic External Defibrillation”, Subcommittee on AED Safety and Efficacy. Circulation, 1997: Vol. 95: 1677-
1682.
b
. Clause 201.7.9.3.103 “Essential Performance data of the Rhythm Recognition Detector” and clause 201.107
“Requirements for Rhythm Recognition Detector,” International Electrotechnical Association, IEC 60601-2-4,
Medical Electrical Equipment – Part 2-4: Particular Requirements for the Basic Safety and Essential Performance
of Cardiac Defibrillators: 2010.
Rhythm category
Requirement
Test result
Shockable (sensitivity):
Coarse VF
Rapid VT
>90%
>75%
Met
Met
Nonshockable (specificity)
>95%
Met
Positive predictive value
Report only
>98%
False positive rate
Report only
<2%
Rhythm category
Minimum sample
size (cases)
Performance
goal
Sample size
tested (cases)
Test result
Shockable (sensitivity):
Coarse VF
Rapid VT
200
50
>90%
>75%
205
80
Met
Met
Nonshockable (specificity):
Normal sinus rhythm
Asystole
Other nonshockable rhythms
300
100
100
30
>99%
>95%
>95%
171
180
385
Met
Met
Met
Intermediate:
Fine VF
Other VT
25
25
Report only
Report only
27
42
66.67% shockable
76.19%
nonshockable
Содержание BeneHeart D2
Страница 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 204: ......
Страница 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Страница 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 207: ......
Страница 208: ...P N 046 010599 00 9 0...