8 - 5
NOTE
•
Total VTBI and Maintain Rate are optional parameters. If the Maintain Rate is
not set, infusion stops at the maintenance stage. If the Total VTBI is not set,
the infusion stops when the syringe is empty.
8.7
Ramp Mode
In the ramp mode, the infusion is running at increasing or decreasing rates.
■
Ramp up stage: in the set ramp up time, the infusion rate increases until steady
rate is reached.
■
Steady stage: the pump infuses at a steady rate.
■
Ramp down stage: in the set ramp down time, the infusion rate decreases until the
set VTBI is completed.
NOTE
•
The Steady Rate can only be obtained by calculation. It is not available for
manual input.
•
Up Time and Down Time are optional parameters. The pump runs an infusion
at the steady rate if they are not set.
8.8
Dose Time Mode
The dose time mode allows the clinician to specify the drug amount, diluent volume or
concentration. The dose mode is typically used for body weight independent drugs.
Rate and time are automatically calculated according to the following formulas:
■
Rate = Dose Rate* Weight/Conc.
■
Time = VTBI/Rate
■
Conc. = Drug Amt. /Volume
Time
Rate
KVO
Total Time
Up Time
Steady
stage
Ramp up stage
Down
Time
Ramp
down
stage
Steady Rate
Содержание BeneFusion eSP Vet
Страница 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Страница 2: ......
Страница 12: ...6 This page intentionally left blank ...
Страница 26: ...2 10 This page intentionally left blank ...
Страница 44: ...5 4 This page intentionally left blank ...
Страница 74: ...11 8 This page intentionally left blank ...
Страница 82: ...12 8 This page intentionally left blank ...
Страница 100: ...A 10 This page intentionally left blank ...
Страница 108: ...B 8 This page intentionally left blank ...
Страница 111: ...D 1 D Declaration of Conformity ...
Страница 112: ...D 2 This page intentionally left blank ...
Страница 113: ......
Страница 114: ...P N 046 021887 00 1 0 ...