M ilw a u ke e
SPECIFICATIONS
SM801 SM802
RANGE
0.0 to 14.0 pH
0.00 to 14.00 pH
0 to 1990 ppm
0 to 4000 ppm
0 to 1990
µ
S/cm 0.00 to 6.00 mS/cm
RESOLUTION
0.1 pH
0.10 pH
10 ppm
10 ppm
10
µ
S/cm
0.01 mS/cm
ACCURACY
±
0.2 pH
±
0.20 pH
(@ 20ºC)
±
2% Full scale
±
2% Full scale
CAL.SOLUTIONS
M10007 (pH 7.01)
M10031 (1413
µ
S/cm)
M10032 (1382 ppm)
M10442 (1500 ppm)
CONVERSION
0.5
0.68
FACTOR
CALIBRATION
Manual, at 1 point
TEMPERATURE COMPENSATION
Automatic, from 0 to 50ºC
PROBE
MA850 combination pH/EC/TDS probe
ENVIRONMENT
0 to 50ºC, 95%RH max
BATTERY TYPE
1x9V alkaline
BATTERY LIFE
approx. 150 hours of use
DIMENSIONS
185 x 82 x 45 mm
WEIGHT
165 g
EC/TDS calibration
The instrument can be manually calibrated at one point
for conductivity range.
A) PREPARATION
For performing TDS calibration,
use 1382 ppm (M10032) solu-
tion for SM801 and 1500 ppm
(M10442) solution for SM802.
For performing EC calibration,
use 1413
µ
S/cm (M10031) so-
lution for both meters.
B) PROCEDURE
Immerse the tip of the probe in the calibration solu-
tion.
Wait for the reading to stabi-
lize, and then turn the EC/TDS
calibration knob until the dis-
play shows the EC or TDS
reading at 25°C.
The calibration is now complete and the meter is ready
for use.
Note: Since the conversion between EC and TDS is
made by a built-in circuit, it is request to
calibrate the meter only in EC or TDS range,
and the other range will be automatically cali-
brated.
The instrument should be recalibrated at least once a
month, or whenever the probe or battery is changed.
BATTERY REPLACEMENT
The meters are powered by a 9V battery that is
located on the rear of the instrument.
When the battery becomes weak, the instrument auto-
matically switches off.
Battery replacement must only take place in a non-
hazardous area using an alkaline 9V battery.
Slide off the battery compartment cover and replace
the rundown battery with a new one. Make sure that
the battery contacts are tight and secure before in-
serting the cover.
pH VERSUS TEMPERATURE CHART
ACCESSORIES
M10004B
pH 4.01 buffer solution, 20
mL sachet (25 pcs)
M10007B
pH 7.01 buffer solution, 20
mL sachet (25 pcs)
M10010B
pH 10.01 buffer solution,
20 mL sachet (25 pcs)
M10031B
1413
µ
S/cm cal. solution,
20 mL sachet (25 pcs)
M10032B 1382 ppm cal. solution, 20
mL sachet (25 pcs)
M10442B 1500 ppm cal. solution, 20
mL sachet (25 pcs)
MA9015
Electrode storage solution,
220 mL bottle
MA850
pH/EC/TDS spare probe
with 1 m cable
ISMIL801 04/03
TEMP
pH VALUES
ºC ºF
0 32
4.01 7.13 10.32
5 41
4.00 7.10 10.24
10 50
4.00 7.07 10.18
15 59
4.00 7.04 10.12
20 68
4.00 7.03 10.06
25 77
4.01 7.01 10.01
30 86
4.02 7.00 9.96
35 95
4.03 6.99 9.92
40 104
4.04 6.98 9.85
45 113
4.05 6.98 9.85
50 122
4.06 6.98 9.82
55 131
4.07 6.98 9.79
60 140
4.09 6.98 9.77
65 149
4.11 6.99 9.76
70 158
4.12 6.99 9.75
M10004B
M10007B
M10010B