M ilw a u ke e

SPECIFICATIONS

   SM801                 SM802

RANGE          

0.0 to 14.0 pH

0.00 to 14.00 pH

         0 to 1990 ppm

0 to 4000 ppm

       0 to 1990 

µ

S/cm      0.00 to 6.00 mS/cm

RESOLUTION 

      

0.1 pH

  0.10  pH

   10 ppm

  10 ppm

   10 

µ

S/cm

  0.01 mS/cm

ACCURACY

  

 

± 

0.2 pH            

 ±

0.20  pH

(@ 20ºC)

                 ±

2% Full scale

                 ±

2% Full scale

CAL.SOLUTIONS          

M10007 (pH 7.01)

                               

M10031 (1413 

µ

S/cm)

               M10032 (1382 ppm)

M10442 (1500 ppm)

CONVERSION           

0.5

   0.68

FACTOR

CALIBRATION           

Manual, at 1 point

TEMPERATURE COMPENSATION

                              Automatic, from 0 to 50ºC

PROBE           

MA850 combination pH/EC/TDS probe

ENVIRONMENT        

0 to 50ºC, 95%RH max

BATTERY TYPE 

            

1x9V alkaline

BATTERY LIFE 

  

    approx. 150 hours of use

DIMENSIONS             

185 x 82 x 45 mm

WEIGHT                          

    165 g

EC/TDS calibration

The instrument can be manually calibrated at one point

for conductivity range.

A)   PREPARATION

For performing TDS calibration,
use 1382 ppm (M10032) solu-
tion for SM801 and 1500 ppm
(M10442) solution for SM802.

For performing EC calibration,
use 1413 

µ

S/cm (M10031) so-

lution for both meters.

B)  PROCEDURE

Immerse the tip of the probe in the calibration solu-
tion.

Wait for the reading to stabi-
lize, and then turn the EC/TDS
calibration knob until the dis-
play shows the EC or TDS
reading at 25°C.

The calibration is now complete and the meter is ready
for use.

Note: Since the conversion between EC and TDS is

made by a built-in circuit, it is request to
calibrate the meter only in EC or TDS range,
and the other range will be automatically cali-
brated.

The instrument should be recalibrated at least once a
month, or whenever the probe or battery is changed.

BATTERY REPLACEMENT

The meters are powered by a 9V battery that is
located on the rear of the instrument.

When the battery becomes weak, the instrument auto-
matically switches off.

Battery replacement must only take place in a non-
hazardous area using an alkaline 9V battery.

Slide off the battery compartment cover and replace
the rundown battery with a new one. Make sure that
the battery contacts are tight and secure before in-
serting the cover.

pH VERSUS TEMPERATURE CHART

ACCESSORIES

M10004B

pH 4.01 buffer solution, 20
mL sachet (25 pcs)

M10007B

pH 7.01 buffer solution, 20
mL sachet (25 pcs)

M10010B

pH 10.01 buffer solution,
20 mL sachet (25 pcs)

M10031B

1413 

µ

S/cm cal. solution,

20 mL sachet (25 pcs)

M10032B              1382 ppm cal. solution, 20

mL sachet (25 pcs)

M10442B              1500 ppm cal. solution, 20

mL sachet (25 pcs)

MA9015

Electrode storage solution,
220 mL bottle

MA850

pH/EC/TDS spare probe
with 1 m cable

ISMIL801   04/03

TEMP

                  pH  VALUES

ºC      ºF

 0        32

    4.01                    7.13                   10.32

 5        41

    4.00                    7.10                   10.24

10      50

    4.00                    7.07                   10.18

15      59

    4.00                    7.04                   10.12

20      68

    4.00                    7.03                   10.06

25      77

    4.01                    7.01                   10.01

30      86

    4.02                    7.00                    9.96

35      95

    4.03                    6.99                    9.92

40     104

    4.04                    6.98                    9.85

45     113

    4.05                    6.98                    9.85

50     122

    4.06                    6.98                    9.82

55     131

    4.07                    6.98                    9.79

60     140

    4.09                    6.98                    9.77

65     149

    4.11                    6.99                    9.76

70     158

    4.12                    6.99                    9.75

M10004B

M10007B

M10010B