CONDITIONING:

Remove the cap and immerse the 

pH54

 in 

MA9015

 storage solution for 2 hours to activate 

the electrode.

CALIBRATION:

Unscrew the battery compartment cap on the 

top of the meter and perform one of the following 

two-point calibration procedure, according to 

the required application:

Acid calibration for measurements below pH7: 

•  Immerse the electrode in pH7.01 

(

M10007

) buffer solution.

•  Allow the reading to stabilize and 

with a small screwdriver turn the 

“7” calibration trimmer inside the 

battery compartment until the display 

shows “7.01”.

• Rinse the electrode with tap 

water and immerse it in pH4.01 

(

M10004

) buffer solution.

•  Allow the reading to stabilize and with a small 

screwdriver turn the “4” calibration trimmer 

until the display shows “4.01”.

Alkaline calibration for measurements above pH7: 

•  Immerse the electrode in pH7.01 (

M10007

buffer solution. 

• Allow the reading to stabilize and with a 

small screwdriver turn the “7” calibration 

trimmer until the display shows “7.01”.

•  Rinse the electrode with tap water and immerse 

it in pH10.01 (

M10010

) buffer solution.

• Allow the reading to stabilize and with a 

small screwdriver turn the “4” calibration 

trimmer until the display shows “10.01”.

Replace the battery compartment cap.

OPERATION:

•  Remove the protective cap.

•  Turn the 

pH54

 on by pressing the ON key.

•  Immerse the electrode in the solution to be 

tested.

•  Stir gently and wait for the reading 

to stabilize.

• After use, rinse the electrode with 

water to minimize contamination.

• Store the electrode with a few 

drops of storage (

MA9015

) or pH7 

(

M10007

) solution in the protective 

cap.

•  Always replace the protective cap after use.

NEVER USE DISTILLED OR DEIONIZED WATER 

FOR STORAGE PURPOSES.

ELECTRODE REPLACEMENT:

The electrode can be easily replaced in the 

following way:

•  Remove the protective cap.

•  Unscrew the plastic ring on the 

top of the electrode.

• Pull out the 

MA73600

 (or 

MA73600-S

) pH electrode and 

replace it with a new one (align-

ing the colored pin towards the 

LCD side).

•  Make sure the gaskets are in place before 

screwing back the ring. 

BATTERY REPLACEMENT:

The 

pH54

 is going to turn off when the bat-

teries are to weak. To replace the batteries 

unscrew the battery compartment cap and 

replace all three 1.5V batteries while paying 

attention to their polarity. Make sure the gasket 

is in place before screwing back the cap.

Batteries should only be replaced in a nonhaz-

ardous area using the battery type specified 

in this instruction manual.

WARRANTY:

This meter is warranted from all defects in materials and manufac-

turing for a period of 

two years

. The 

electrode is warranted for 

a period of one year. 

During this period, the required repair or 

replacement of parts, where the damage is not due to negligence 

or erroneous operation by the user, will be effected free of charge. 