manualshive.com logo in svg
background image

USER MANUAL 

 

pH 58 

Pocket-size 

pH/ORP/°C/°F Meter 

 

 

 

 

www.milwaukeeinst.com 

 

 

Automatic Temperature 

Measuring  unit 

Compansation Indicator 

on main LCD 

Instability Indicator 

Main LCD 

 
Calibration Indicator 

Temperature 

Low Battery Indicator 

reading  

 

Secondary LCD 

 
 
OPERATIONAL GUIDE 
• Remove the electrode cap and immerse the 
meter in 

MA9015 

storage solution for 2 hours 

to activate the electrode. 
• Turn the meter on by pressing the ON/OFF 
button. All the used segments on the LCD will 
be visible for 1 second or as long as the button 
is pressed. 
• Immerse the electrode in the solution to be 
tested. Stir gently and wait for the reading to 
stabilize, i.e. the instability indicator 
(hourglass) on the LCD turns off. 
• Select the pH or ORP (mV) mode by 
pressing the SET/HOLD button. 
• The displayed pH value is automatically 
compensated for temperature. The ORP (mV) 
value will be displayed on the main LCD, while 
the temperature will always be displayed on 
the secondary LCD. 
• 

To freeze the display

, while in measurement 

mode, press and hold the SET/HOLD button. 
The “HOLD” message appears on the 
secondary display and the reading will be 
frozen on the LCD. Press any button to return 
to the normal measurement mode. 
• 

To turn the meter off

, press the ON/OFF 

button. The “OFF” message will appear on 
the secondary display. Release the button. 

Note: 

• Before taking any measurement, make sure 
the meter is calibrated (the CAL tag is on). 
• After use always turn the meter OFF, rinse 
the electrode with water to minimize 
contamination  and store it with a few drops of 
storage (

MA9015

) or pH7 (

M10007

) solution in 

the protective cap. 
 
NEVER USE DISTILLED OR DEIONIZED 
WATER FOR STORAGE PURPOSE. 
 
CALIBRATION PROCEDURE 

 
The ORP (mV) range is factory calibrated

Contact the nearest Service Center for 
recalibration, if necessary. 
For better accuracy, frequent calibration of the 
tester is recommended. Calibration is also 
necessary  after electrode replacement, after 
testing  aggressive chemicals and where 
extreme accuracy is required. 
• Turn the meter on and select the pH mode 
with the SET/HOLD button. 
• Press and hold the ON/OFF/CAL button until 
the “OFF” message on the secondary LCD  is 
substituted by “CAL”. Release the button. 
• The instrument enters the calibration mode 
by displaying “pH 7.01 USE” (or “pH 6.86 
USE” if the NIST buffer set was selected). 
• For a 

single-point calibration

, immerse the 

electrode in any buffer, i.e. pH 4.01, 7.01 (or 
6.86), 10.01 (or 9.18).  
• The meter activates the automatic buffer 
recognition.  If no valid buffer is detected, the 

meter keeps the USE indication active for 12 
seconds, and then replaces it with WRNG, 
indicating that the sample being measured is 
not a valid buffer. Otherwise, if a valid buffer is 
detected, then  its value is shown on the 
primary display, and REC appears on the 
secondary LCD. 
• If the pH 7.01 (or pH 6.86) was used, press 
the SET button to exit the Calibration mode 
and the “OK 1” message will appear on the 
display. The first calibration point is stored and 
the meter returns to normal measurement 
mode. 
 

For better accuracy, it is always 
recommended 

to perform a 2-point 

calibration. 
 

• For a 

two-point calibration

, immerse the 

electrode  in pH 7.01 (or pH 6.86) buffer 
solution. 
• After the first point has been accepted, the 
meter will then ask for the second buffer and 
the message “pH 4.01 USE” will appear. 
• Rinse the electrode and immerse it in the 
second solution (pH 4.01, 10.01 or 9.18). 
• If a valid buffer value is detected, the REC 
message is displayed and the meter 
completes the calibration procedure. The LCD 
shows the accepted value with the “OK 2” 
message and the instrument returns to the 
normal measurement mode. Otherwise, if no 
valid buffer is detected, the meter displays the 
WRNG message. 

 

Отзывы: