Dear Customer,

Thank you for choosing a Milwaukee Instruments 

product. This manual will provide you with the necessary 

information for the correct use of the instrument. Please 

read it carefully before using the meter. 

SPECIFICATIONS

Range

0.00 to 10.00 mg/L Cl

2

Resolution

0.01 mg/L (0.00-3.50 mg/L)

0.10 mg/L (above 3.50 mg/L)

Precision

±0.17 mg/L @5.00 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell and 525 nm narrow 

band interference filter

Method

Adaptation of USEPA method 330.5 and 

Standard method 4500 - Cl G

Environment

0 to 50 °C (32 to 122 °F) 

100% RH max

Battery Type

9 volt (1 pc)

Auto-Shut off

After 10’ of non-use

Dimensions

192 x 104 x 52 mm 

(7.5 x 4.1 x 2”)

Weight

380 g

MEASUREMENT PROCEDURE

1.  Turn the meter on by pressing ON/OFF.

2.  When the LCD displays “- - -”, the 

meter is ready.

3.  Fill the cuvet with 10 mL of sample, up 

to the mark, and replace the cap.

4.  Place the cuvet into the holder and 

ensure that the notch on the cap is 

positioned securely into the groove.

5.  Press ZERO and “SIP” will blink on the 

display.

6.  After a few seconds the display will 

show “-0.0-”. The meter is now zeroed 

and ready for measurement.

7.  For free chlorine measurement:  

Add 3 drops of DPD 1 reagent to an empty cuvet.

 

 

8.  Add 5 mL of DPD 0 reagent by using a syringe.

9.  Fill the cuvet up to the 10 mL mark with 5 mL of 

unreacted sample using a pipette.

10. Replace the cap and shake gently for a few seconds.

DPD 1

x 3

11.  Immediately insert the cuvet into the holder and 

ensure that the notch on the cap is positioned 

securely into the groove.

12. Press READ and “SIP” will blink 

during measurement.

13. The instrument directly displays 

concentration in mg/L of chlorine.

14. Then, for total chlorine measurement, remove the 

cuvet and add to the reacted sample 3 drops of DPD 

3. Replace the cap and swirl gently.

DPD 3

x 3

 

15. Reinsert the cuvet into the holder and 

ensure that the notch on the cap is 

positioned securely into the groove.

16. Wait for 2 minutes and 30 seconds 

then press READ and “SIP” will blink 

during measurement.

17. The instrument directly displays concentration in 

mg/L of total chlorine.

Interferences

 

y

Bromine (positive error)

 

y

Chlorine dioxide (positive error)

 

y

Iodine (positive error)

 

y

Oxidized Manganese and Chromium (positive error)

 

y

Ozone (positive error)

 

y

Alkalinity above 250 mg/L CaCO3 or acidity above 150 

mg/L CaCO3 will not reliable develop the full amount of 

color or it may rapidly fade. To resolve this, neutralize 

the sample with diluted HCl or NaOH.

2’30”

GUIDE TO DISPLAY CODES

This prompt appears for 1 second 

each time the instrument is turned 

on.
The dashes “- - -” indicates that the 

meter is in a ready state and zeroing 

can be performed.
Sampling In Progress. Flashing “SIP” 

prompt appears each time the meter 

is performing a measurement.

“-0.0-”, the meter is in a zeroed state 

and measurement can be performed.

The blinking “BAT” indicates that the 

battery voltage is getting low and the 

battery needs to be replaced.
“-bA-”, the battery is dead and must 

be replaced. Once this indication 

is displayed, the meter will lock up. 

Change the battery and restart the 

meter.
“Conf”, the meter has lost its 

configuration. Contact your dealer 

or the nearest Milwaukee Customer 

Service Center.

ERROR MESSAGES

On zero reading

 

y

Blinking “-0.0-” indicates that the 

zeroing procedure failed due to a 

low signal-to-noise ratio. In this 

case press ZERO again.

 

y

“no L”, the instrument can not 

adjust the light level. Please check 

that the sample does not contain 

any debris.

 

y

“L Lo”, there is not enough light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.

 

y

“L Hi”, there is too much light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.

10 ml