SPECIFICATIONS

Free & Total Chlorine

Range

0.00 to 5.00 mg/L Cl

2

Resolution

0.01 mg/L (0.00-3.50 mg/L);

0.10 mg/L (above 3.50 mg/L)

Precision

±0.06 mg/L @ 1.50 mg/L

Method

Adaptation of the USEPA method 330.5 

and Standard Method 4500-Cl G.

pH

Range

6.5 to 8.0 pH

Resolution

0.1 pH

Precision

±0.1 pH @ 7.2 pH

Method

Adaptation of the phenol red method.

Other

Light Source

Tungsten lamp

Light Detector

Silicon Photocell and 525 nm narrow 

band interference filter

Environment

0 to 50 °C (32 to 122 °F) 

100% RH max

Battery Type

9 volt (1 pc)

Auto-Shut off

After 10’ of non-use

Dimensions

192 x 104 x 52 mm 

(7.5 x 4.1 x 2”)

Weight

380 g

CHANGE PARAMETER

To change parameter between Free & 

Total Chlorine and pH, press and hold for 3 

seconds, the ZERO key.

The selected parameter will be displayed as 

“Cl” (chlorine) or “PH” on the display.

 or 

MEASUREMENT PROCEDURE

1.  Turn the meter on by pressing ON/OFF. The last selected 

parameter is displayed on the Liquid Crystal Display 

(“Cl” or “PH”).

2.  Choose the parameter that you want to 

measure.

3.  Fill the cuvet with 10 mL of sample, up to 

the mark, and replace the cap.

4.  Place the cuvet into the holder and ensure 

that the notch on the cap is positioned 

securely into the groove.

5.  Press ZERO and “SIP” will blink on the 

display.

6.  After a few seconds the display will show 

“-0.0-”. The meter is now zeroed and ready 

for measurement.

 

 

For Free Chlorine measurement

7.  Add 3 drops of DPD 1 reagent and 3 drops of DPD 2 

reagent to an empty cuvet.

8.  8• Immediately add 10 mL of unreacted sample. Replace 

the cap and shake gently.

 

10 mL

 ▶

DPD 1

x 3

DPD 1

x 3

+

9.  Immediately reinsert the cuvet into the holder and 

ensure that the notch on the cap is positioned securely 

on the groove.

10.  Press READ and “SIP“ will blink during measurement.

 

 

11.  The instrument directly displays concentration in mg/L 

of free chlorine.

For Total Chlorine measurement

12.  Remove the the cuvet and add to the 

reacted sample 3 drops of DPD 3 reagent. 

Replace the cap and swirl gently.

13.  Reinsert the cuvet into the holder and ensure that the 

notch on the cap is positioned securely into the groove.

14.  Wait for 2 minutes 30 seconds then press READ and 

“SIP” will blink during measurement.

10 ml

DPD 3

x 3

2’30”

 

 

 

15.  The instrument dirrectly displays concentration in mg/L 

of total chlorine.

For pH measurement

16.  Fill a cuvet with 10 mL of unreacted sample 

up to the mark.

17.  Add 5 drops of pH-0 reagent. Replace the 

cap and shake gently.

18.  Insert the cuvet into the holder and ensure that the 

notch of the cap is positioned securely into the groove.

19.  Press READ and “SIP” will blink during measurement.

20. The instrument dirrectly display the pH value.

GUIDE TO DISPLAY CODES

This prompt appears for 1 second each 

time the instrument is turned on.

The dashes “- - -” indicates that the 

meter is in a ready state and zeroing 

can be performed.

Sampling In Progress. Flashing “SIP” 

prompt appears each time the meter is 

performing a measurement.

“-0.0-”, the meter is in a zeroed state 

and measurement can be performed.

The blinking “BAT” indicates that the 

battery voltage is getting low and the 

battery needs to be replaced.
“-bA-”, the battery is dead and must 

be replaced. Once this indication is 

displayed, the meter will lock up. Change 

the battery and restart the meter.
“Conf”, the meter has lost its configuration. 

Contact your dealer or the nearest 

Milwaukee Customer Service Center.

ERROR MESSAGES

On zero reading

 

y

Blinking “-0.0-” indicates that the zeroing 

procedure failed due to a low signal-to-

noise ratio. In this case press ZERO again.

pH - 0

x 5