background image

Dear Customer,

Thank you for choosing a Milwaukee Instruments 

product. This manual will provide you with the necessary 

information for the correct use of the instrument. Please 

read it carefully before using the meter. 

SPECIFICATIONS

Range

0.00 to 3.00 mg/L (NH

- N)

Resolution

0.01 mg/L

Precision

±0.09 mg/L @1.50 mg/L

Light Source

Blue LED 466 nm

Light Detector

Silicon Photocell and 466 nm narrow 

band interference filter

Method

Adaptation of Nessler method

Environment

0 to 50 °C (32 to 122 °F) 

100% RH max

Battery Type

9 volt (1 pc)

Auto-Shut off

After 10’ of non-use

Dimensions

192 x 104 x 52 mm 

(7.5 x 4.1 x 2”)

Weight

380 g

MEASUREMENT PROCEDURE

1.  Turn the meter on by pressing ON/OFF.

2.  When the LCD displays “- - -”, the 

meter is ready.

3.  Fill the cuvet with 10 mL of sample, up 

to the mark, and replace the cap.

4.  Place the cuvet into the holder and 

ensure that the notch on the cap is 

positioned securely into the groove.

5.  Press ZERO and “SIP” will blink on the 

display.

6.  After a few seconds the display will 

show “-0.0-”. The meter is now zeroed 

and ready for measurement.

7.  Remove the cuvet. Open it.

 

 

8.  Add 4 drops of NH

3

-1 (10 drops in case of seawater 

analysis) and swirl gently.

9.  Add 4 drops of NH

3

-2 (10 drops in case of seawater 

analysis). Replace the cap and swirl gently.

 

10 mL

 ▶

NH3 1

NH3 2

x 4

x 4

+

10. Immediately reinsert the cuvet into 

the holder and ensure that the notch 

on the cap is positioned securely into 

the groove.

11.  Press and hold READ for 3 seconds 

and the display will show the 

countdown prior to the measurement. 

Alternatively, wait for 3 minutes and 

30 seconds and just press READ. In 

both cases “SIP” will blink during 

measurement.

12. The instrument directly displays the concentration in 

mg/L of ammonia nitrogen (NH

3

-N).

13. To convert the reading to mg/L of ammonia (NH

3

), 

multiply by a factor of 1.216.

Interferences

 

y

Chloramines, various aliphatic and aromatic 

amines, glycine, (positive error). To eliminate these 

interferences distillation is required.

 

y

Aldehydes, alcohols (e.g. ethanol) or acetone (negative 

error). To eliminate these interferences distillation is 

required.

 

y

Sulfide (S

2-

): may cause turbidity.

 

y

Hardness above 1 g/L as Calcium Carbonate (CaCO

3

).

GUIDE TO DISPLAY CODES

This prompt appears for 1 second 

each time the instrument is turned 

on.
The dashes “- - -” indicates that the 

meter is in a ready state and zeroing 

can be performed.
Sampling In Progress. Flashing “SIP” 

prompt appears each time the meter 

is performing a measurement.

3’30”

“-0.0-”, the meter is in a zeroed state 

and measurement can be performed.

The blinking “BAT” indicates that the 

battery voltage is getting low and the 

battery needs to be replaced.
“-bA-”, the battery is dead and must 

be replaced. Once this indication 

is displayed, the meter will lock up. 

Change the battery and restart the 

meter.
“Conf”, the meter has lost its 

configuration. Contact your dealer 

or the nearest Milwaukee Customer 

Service Center.

ERROR MESSAGES

On zero reading

 

y

Blinking “-0.0-” indicates that the 

zeroing procedure failed due to a 

low signal-to-noise ratio. In this 

case press ZERO again.

 

y

“no L”, the instrument can not 

adjust the light level. Please check 

that the sample does not contain 

any debris.

 

y

“L Lo”, there is not enough light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.

 

y

“L Hi”, there is too much light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.

On sample reading 

 

y

“-SA-”, there is too much light for 

the sample measurement. Please 

check if the right sample cuvet is 

inserted.

 

y

“Inv”, the sample and the zero 

cuvet are inverted. 

10 ml

Отзывы: