10

Instruction Manual Mi170 Bench Meter

EC/TDS  CALIBRATION

EC calibration is a one-point procedure. Selectable calibration points are 0.00 µS for 
offset and 84.0 µS, 1413 µS, 5.00 mS, 12.88 mS, 80.0 mS, 111.8 mS for slope.
Rinse the probe with calibration solution or deionized water. Immerse the probe into 
the solution. The sleeve holes must be completely submerged. Tap the probe repeatedly 
to remove any air bubbles that may be trapped inside the sleeve.
To enter EC calibration, select the EC range and press CAL.

Note

: The TDS reading is automatically derived from the EC reading and no specific 

calibration  for  TDS  is  needed.  Pressing  CAL  when  TDS  range  is  selected  has 
no effect.

For 

zero calibration

, just leave the dry probe in the air. This calibration is performed in 

order to correct the reading around 0.00 µS. The slope is evaluated when the calibration 
is performed in any other point.
The  primary  LCD  will  display  the  EC  reading  and  the  secondary  LCD  will  display  the 
closest calibration solution, along with “CALIBRATION” tag. The “WAIT” tag will blink 
until the reading is stable.

When  the  reading  is  stable  and  close  to  the  selected  calibration  solution,  “READY” 
and “ACCEPT” tags will blink on the LCD.

23

www.milwaukeeinst.com

SETUP

Setup mode allows viewing and modifying the following parameters:

•  Cell Constant

•  TDS Factor

•  Temperature Coefficient

•  Reference Temperature

•  Current Time (hh:mm)

•  Current Date (MM.DD.YYYY)

•  Beep Status

•  Baud Rate (serial communication)

•  Instrument ID

•  Temperature Unit
To  enter  SETUP  mode,  press  and  hold  the  SETUP  key  for  about  2  seconds  while  in 
normal measurement mode.

Select the desired setup parameter using the UP and DOWN arrow keys.

Press the CAL key if you want to change the item value. The selected item (e.g. hour, 
in setting up the correct time) and “ACCEPT” tag will start blinking.

Press the arrow

 

keys to change the displayed value.

If there is another item to be set (e.g. minutes), press the RANGE/FIXED key. The other 
item will start blinking.

Содержание Mi170

Страница 1: ...sor 11 H 6726 Szeged HUNGARY tel 36 62 428 050 fax 36 62 428 051 www www milwaukeeinst com e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA Tel 1 252 443 3630 Fax 1 252 443 1937 e mail sales milwaukeetesters com 1 www milwaukeeinst com www milwaukeeinst com INSTRUCTION MANUAL Bench Meter Mi170 EC TDS NaCl Temperature ...

Страница 2: ...DESCRIPTION 31 www milwaukeeinst com WARRANTY This instrument is warranted against defects in materials and manufacturing for a period of 3 years from the date of purchase Electrodes are warranted for 6 months If during this period the repair or replacement of parts is required where the damage is not due to negligence or erroneous operation by the user please return the intrument electrode and pr...

Страница 3: ...ter fails to calibrate NaCl Incorrect EC calibration Recalibrate the meter in EC range Set cell constant to 1 At startup the meter displays all LCD tags permanently One of the keys is blocked Check the keyboard or con tact your dealer Err xx message displayed at start up Internal error Contact your dealer or any Milwaukee Instruments Service Center 3 www milwaukeeinst com FUNCTIONAL DESCRIPTION FR...

Страница 4: ... temperature coefficient is user selectable Mi170 includes also GLP capability and data transfer to a computer through a RS232 or a USB port In addition the meter allows the user to enter an ID code to uniquely identify the instrument For accurate measurements use the electrode holder supplied with the meter This Bench Meter is supplied with MA814DB 1 EC Temperature probe MA9315 Electrode Holder M...

Страница 5: ...amples 12 chars LDEC Requests the xxxth EC record logged data LDTD Requests the xxxth TDS record logged data LDNC Requests the xxxth NaCl record logged data LAEC Requests all EC Log on demand LATD Requests all TDS Log on demand LANC Requests all NaCl Log on demand Notes Err8 is sent if the instrument is not in measurement mode Err6 is sent if the requested range is not available Err4 is sent if th...

Страница 6: ...y sis To connect the instrument to a PC through the RS232 port use the Milwaukee MA9350 cable connector To connect the instrument to a PC through the USB port use a standard USB cable Make sure that your instrument is switched off and plug one connector of the cable to the instrument RS232 USB connector and the other to the serial port of your PC Notes Other cables than MA9350 may use a different ...

Страница 7: ... 115700 117700 119700 121800 123900 2760 3180 3615 4063 4155 4245 4337 4429 4523 4617 4711 4805 4902 5000 5096 5190 5286 5383 5479 5575 7 www milwaukeeinst com OPERATIONAL GUIDE INITIAL PREPARATION Plug the 12 VDC adapter into the power supply socket Connect the EC TDS probe to the 7 pin connector Make sure the probe sleeve is properly inserted and tighten the threaded ring Turn the instrument on ...

Страница 8: ...nst com TEMPERATURE CALIBRATION for technical personnel only The Mi170 is factory calibrated for temperature Milwaukee s temperature probes are interchangeable and no temperature calibration is needed when they are replaced If the temperature measurements are inaccurate temperature recalibration should be performed For an accurate recalibration contact your dealer or the nearest Milwaukee Customer...

Страница 9: ... ATC mode or 400 0 for uncompensated TDS will be displayed blinking If the SHIFT RANGE FIXED keys are pressed to freeze the LCD range and the reading goes out of range the full scale value of the frozen range will be displayed blink ing NaCl MEASUREMENTS Press the RANGE FIXED key to enter NaCl measurement mode The instrument will display the NaCl reading on the primary LCD and the temperature on t...

Страница 10: ... point The primary LCD will display the EC reading and the secondary LCD will display the closest calibration solution along with CALIBRATION tag The WAIT tag will blink until the reading is stable When the reading is stable and close to the selected calibration solution READY and ACCEPT tags will blink on the LCD 23 www milwaukeeinst com SETUP Setup mode allows viewing and modifying the following...

Страница 11: ...k Calibration can not be confirmed In this case check if the correct calibration solution has been used If the meter is in ATC mode and the solution temperature is out of the 0 0 to 60 0 C 32 0 to 140 0 F interval the WRONG SOLUTION C tags and the temperature will be displayed blinking For best results choose an EC calibration solution value close to the sample to be measured It is possible to set...

Страница 12: ...xpected value the WRONG and SOLUTION tags will blink Calibration cannot be confirmed If the temperature of the buffer is out of the 0 0 60 0 C 32 0 140 0 F temperature interval the WRONG SOLUTION and C tags will be displayed blinking If a new EC calibration is performed the NaCl calibration is automatically cleared Thus a new NaCl calibration is required 21 www milwaukeeinst com or For TDS range t...

Страница 13: ... storage and retrieval of data regarding the main tenance and status of the electrode All data regarding EC and NaCl calibration is stored for the user to review when necessary LAST EC CALIBRATION DATA Last EC calibration data is stored automatically after a successful calibration To view the EC calibration data press the GLP ACCEPT key when the instrument is in EC measurement mode The instrument ...

Страница 14: ...ure com pensation mode 19 www milwaukeeinst com Otherwise the instrument will display the EC TDS or NaCl value on the primary LCD and the temperature on the secondary LCD along with last stored record number and LOG tag Pressing the arrow keys the instrument will display the same parameter but for a dif ferent record Press the RANGE FIXED key and the instrument will display the next logged paramet...

Страница 15: ... measure ment mode If no data were logged the instrument displays no rEC message for the selected range 15 www milwaukeeinst com The instrument ID LAST NaCl CALIBRATION DATA Last NaCl calibration data is stored automatically after a successful calibration To view the NaCl calibration data press the GLP ACCEPT key when the instrument is in NaCl measurement mode The instrument will display the time ...

Страница 16: ...ange the instrument displays no CAL message blinking 17 www milwaukeeinst com LOGGING Up to 50 LOG samples can be stored into memory for each measurement range EC TDS and NaCl LOGGING THE CURRENT DATA To store the current reading into memory press the LOG MR key while in measure ment mode The instrument will display the current date MM DD on the primary LCD and the record number on the secondary L...

Отзывы: