Free & Total Chlorine 

High Range Meter

 

 

 

 

 

 

 

 

 

 

Dear Customer, 
Thank you for choosing a Martini product. This 
manual will provide you with the necessary 
information for the correct use of the instrument.  
Please read it carefully before using the meter. 

This instrument is in compliance with 

 

directives. 

 

SPECIFICATIONS:

 

Range 

0.00 to 10.00 mg/L Cl2 

Resolution

 

0.01 mg/L (0.00-3.50 
mg/L);0.10 mg/L (above 
3.50 mg/L) 

Precision

 

±0.10 mg/L @ 5.00 mg/L 

Light Source

 

Tungsten lamp 

Light 

Detector

 

Silicon Photocell and 525 
nm narrow band 
interference filter 

Method

 

Adaptation of the USEPA 
method 330.5 and 
Standard Method 4500-
Cl G. 

Environment 

0 to 50°C (32 to 122°F); 
100% RH max. 

Battery Type 

1 x 9 volt 

Auto-Shut off 

After 10' of non-use 

Dimensions 

192 x 104 x 52 mm 
(7.5 x 4.1 x 2”) 

Weight 

380 g 

 
 
 
 
 
 

MEASUREMENT PROCEDURE: 
 

1•   Turn the meter on by pressing 

ON/OFF. 

2•   When the LCD displays “-  -  -”,  the 

meter is ready. 

3•  

Add 5 mL of DPD 0 
reagent by using a 
syringe. 

4•  

Fill the cuvet up to the 
10 mL mark with 5 mL 
of unreacted sample 
using a pipette. 

5•

  Replace the cap  and 

shake  gently  for a few 
seconds.

 

6• 

Reinsert the cuvet into 
the holder and ensure 
that the notch on the 
cap is positioned 
securely into the 
groove.

 

7•   Press ZERO and “SIP” will blink on the 

display. 

8•   After a few seconds the  display will 

show “-0.0-”. The meter is now zeroed 
and ready for measurement. 

 

 

 
9• 

Remove the cuvet. Open it. 

10• 

For free chlorine measurement

:  

 

Add 3 drops of DPD 1 reagent. 

 

Replace the cap and swirl gently. 

  

 

 

11•   Immediately  insert the cuvet into the 

holder  and ensure that the notch on 

the cap is positioned securely into the 
groove. 

12•   Press READ and “SIP” 

will blink 

during 

measurement. 

13•   The instrument directly 

displays concentration in 
mg/L of chlorine. 

14•  Then, 

for total chlorine measurement

remove the cuvet and add to the 
reacted  sample 3 drops of DPD 3. 
Replace the cap and swirl gently. 

 

15•   Reinsert the cuvet into the holder and 

ensure  that the notch on the cap is 
positioned securely into the groove. 

16•   Wait for 2 minutes and  30 seconds 

then press 
READ and 
“SIP” will blink 
during 
measurement. 

17•   The instrument directly displays 

concentration 

in mg/L of total 

chlorine. 

 

Interferences 
 

• 

 Bromine (positive error) 

•  

Chlorine dioxide (positive error) 

•  

Iodine (positive error) 

•  

Oxidized Manganese and Chromium 
(positive error) 

•  

Ozone (positive error) 

•  

Alkalinity above 250 mg/L CaCO3 or 
acidity above 150 mg/L CaCO3 will not 
reliable develop  the full amount of 
color or it may rapidly fade. To resolve 
this, neutralize the sample with diluted 
HCl or NaOH.

 