USER MANUAL 

 

MI 405 

Ammonia Medium 

Range Meter 

 

 

 

 

 

 

 

 

 

 

Dear Customer, 
Thank you for choosing a Martini product. 
This manual will provide you with the 
necessary information for the correct use of 
the instrument.  
Please read it carefully before using the 
meter. This instrument is in compliance with 

 directives. 

 

 

SPECIFICATIONS:

 

Range

  

0.00 to 9.99 mg/L (NH

3

-

N)  

Resolution

   0.01 mg/L 

 

Precision

  

±0.10 mg/L @ 5.00 mg/L   

Light Source 

Blue LED 466 nm 

 

Light Detector

  Silicon Photocell and 

466 nm narrow band interference filter   

Method

 Adaptation of Nessler method   

Environment

 0 to 50°C (32 to 122°F); 

100% RH max. 

 

Battery Type

 1 x 9 volt 

 

Auto-Shut off

 After 10' of non-use 

 

Dimensions  

192 x 104 x 52 mm 

(7.5 x 4.1 x 2”) 

 

Weight

  

380 g 

 

 

 
 
 
 
 
 
 
 
 

MEASUREMENT PROCEDURE: 
 

1•   Turn the meter on by pressing 

ON/OFF. 

2•   When the LCD displays “-  -  -”,  the 

meter is ready. 

3•   Fill the cuvet with 10 

mL of sample, up to the 
mark, and  replace the 
cap. 

4•   Place the cuvet into the 

holder  and ensure that 
the notch on the cap is 
positioned securely into 
the groove. 

5•   Press ZERO and “SIP” 

will blink on the display. 

6•   After a few seconds the 

display will show “-0.0-
”. The  meter is now 
zeroed and ready  for 
measurement. 

 

 

 
7•   Remove the cuvet. Open it. 
8•   Add 4 drops of NH3-1 (10 drops in 

case of seawater analysis) and swirl 
gently. 

9•   Add 4 drops of NH3-2 (10 drops in 

case of  seawater analysis). Replace 
the cap and swirl gently. 

 

 

 

10• Immediately reinsert 

the cuvet  into the 
holder and ensure 
that the notch on the 
cap is positioned 
securely into the 
groove. 

11• Press and hold 

READ for 3 seconds 
and the display will 
show the countdown 
prior to 

the 

measurement. 
Alternatively, wait 
for 3 minutes and 30 
seconds and just press  READ. In 
both cases “SIP” will  blink during 
measurement. 

12• The instrument directly displays the 

concentration  in mg/L of ammonia 
nitrogen (NH3-N). 

13• To convert the reading to mg/L of 

ammonia (NH3), multiply by a factor 
of 1.216. 

 
Interferences 
•   Organic compounds like: 

chloramines, various  aliphatic  and 
aromatic amines, glycine,  (positive 
error). 
To eliminate these interferences 
distillation is required. 

•   Organic compounds like: aldehydes, 

alcohols  (e.g. ethanol) or acetone 
(negative error). 
To eliminate these interferences 
distillation is required. 

•   Sulfide (S2-): may cause turbidity. 
•   Hardness above 1 g/L as Calcium 

Carbonate (CaCO3). 