USER MANUAL 

 

MI 106 

pH/mV/Temperature 

Portable Meter 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 

GENERAL OPERATION 
 

•   The meter is supplied 
complete 

with a 9V 

battery. 

Remove the 

battery compartment 
cover on the  back of the 
meter and install  the 
battery while paying 
attention to its polarity. 
•   Connect the 

MA923D/1 

probe to the 

meter. 
•   Turn the instrument 
on by 

pressing the 

ON/OFF key. At start-up, 
the LCD 
shows the percentage of  the remaining 
battery life  for a few seconds, and  then 
the current measurement. 
•    Always remove the electrode 
protective cap 

before taking any 

measurement. If the electrode  has been 
left dry, soak the tip (bottom  2.5 cm) in 

M10000 

rinse solution for a few minutes. 

• 

 The temperature reading can be 

displayed  in °C or °F unit; to select the 
desired scale,  press and hold the 
“ON/OFF” key until “TEMP”  and the 
current temperature unit are displayed 
on the secondary LCD. Use the “SET” 
button to select the unit and then press 
the  “ON/OFF” key a couple of times to 
return to normal mode. 
•   To activate the HOLD function,  keep 
pressed  the “HOLD” key. The measured 
value will  be frozen on the display and 
the “HOLD”  message appear on the 
secondary LCD. 
 

 

 

•   After measurements, switch the 
meter off  by pressing the “ON/OFF” key. 
The “OFF”  message appears on the 
LCD; release the button. 
• 

After use, replace the electrode 

protective cap with a few drops of 
storage solution. 
 

TAKING pH MEASUREMENTS 
 

•   Select the “pH” mode with the “SET ” 
key. 
•   Immerse the tip (2.5 cm) of the probe 
in the sample and stir gently. 
•   Read the pH value when the clock 
symbol stops blinking. 

Note: 

Before taking any measurement, 

make sure that the meter has been 
calibrated (the “CAL” tag is displayed on 
the left lower corner of the LCD). 

pH CALIBRATION PROCEDURE 

It is recommended to recalibrate the 
meter at least once every three weeks. 
A) Preparation, Buffer solutions: 
1. 

pH 7.01 

6.86 (MA9007 / MA9006) 

2. 

pH 4.01 

(

MA9004

) for acidic 

calibration 
(pH < 7) or 

pH 10.01 / 9.18 

(

MA9010 / 

MA9009

) for alkaline range (pH > 7). 

Use two beakers for each buffer solution: 
one beaker for rinsing the probe, the 
other for calibration. In this way the cross 
contamination between solutions is 
minimized. 
B) Procedure: 
•   Select the “pH” mode with the “SET ” 
key. 
•   Select the calibration buffer set: 
press and hold the “ON/OFF” key until 
the LCD shows “TEMP”. Press again this 
key and the “BUFF” message will 
appear; then select the desired buffer set 
with the “SET” key: “7.01 pH BUFF” (for 
standard solutions: pH 4.01, 7.01, 10.01) 
or “6.86 pH BUFF” (for NIST solutions: 

pH 

4.01, 6.86, 9.18). Press the 

“ON/OFF” key again to exit. 
 
•    Remove the protective cap, then 
immerse  the probe in the first buffer (pH 
7.01 or 6.86). 
•    Press and hold the “ON/OFF” key 
until “CAL”  is displayed on the lower 
LCD. 
 
 
 
 
 
•    Release the button 
and the  message “7.01 
pH USE” (or  “6.86 pH 
USE” for NIST buffer set) 
will be displayed. 
•    The meter automatically recognizes 
the buffer  value: if a valid buffer is 
detected, then its  value is displayed on 
the LCD  together with 
the “REC” message, 
and the first calibration 
point is accepted. 
•    The meter will then 
require 

the second 

buffer by showing  the 
“pH 4.01 USE” 
message.  
Immerse the probe in  the second buffer 
solution  (pH 4.01 or 10.01 or 9.18). If a 
valid buffer  is recognized, the meter 
completes the calibration  procedure, 
shows the “OK 2” message  for a few 
seconds, then returns to  the normal 
mode. 
 

Notes: 

•   If no valid buffer is 
detected, the LCD will 
show the “WRNG” 
message. 
•   For exiting the calibration after the 
first point is accepted,  press the 