OPERATION: 

• 

Connect  the  supplied  12  VDC

power  adapter  to  the  meter  and
to the mains line.

• 

Connect  the  pH  and  ORP  elec-

trode  to  the  BNC  socket  of  the
meter.

• 

Always  remove  the  electrode

protective caps  before taking  any
measurement.  If  the  electrodes
have  been  left  dry,  soak  the  tip
(bottom  2.5  cm)  in  rinse  solution
(M10000B) for a few minutes to reactivate them.

• 

Make  sure  that  the  meter  has  been  calibrated  before  taking  any

measurements (see Calibration Procedure).

• 

Immerse the tip (2.5 cm) of the electrodes into the samples.

• 

Turn the instrument on by pressing the ON/OFF key.

• 

Allow  the  reading  to  stabilize  and  the  meter  will  start  continuous

monitoring.

• 

A  blinking  alarm  LED  will  indicate  when  the  measured  pH  value  is

higher or lower than the setpoint depending on user selection.

• 

When  the  alarm  LED is  blinking, there  is  power  in the  output  power

socket.

NOTE: 

  The  output  power  contact  has  no  protection  fuse  inside  the 

meter. It is recommended to protected it outside, against failure. 

SETPOINT:

 

• 

The  setpoint  can  be  selected  by  adjusting  the  knob  to  the  desired
value.

• 

The  selectable range for  MC125  is  from
4  to  8  for  pH  and  from  -200  to  600  mV
for ORP.

• 

The  nature  of  the  setpoint  can  be  se-
lected by setting the proper switch to the
desired position (ABOVE or BELOW).

CALIBRATION PROCEDURE:

 

• 

Remove the protective cap from the pH electrode.

• 

Immerse  the  electrode  tip  into  a  new  sachet  of  pH  7
calibration solution and allow the reading to stabilize.

• 

Adjust  the  pH7  calibration  trimmer  (on  the  front)  to
display "7.00 pH".

• 

Open  a  new  sachet  of  pH  4  calibration  solution  and
use a small quantity to rinse the electrode.

• 

Immerse  the  electrode  tip  inside  the  pH  4  sachet  and
allow the reading to stabilize.

• 

Adjust  the  pH4/10  calibration  trimmer  (on  the  front)  to
display "4.00 pH".

• 

The calibration is now complete and the meter is ready
to take measurements.

• 

It  is  suggested  to  recalibrate  the  meter  at  least
once  a  month,  after  prolonged  warehousing  time
and after pH electrode replacement.

INSTALLATION PROCEDURE: 