XRS-5 Transducer
PL-590
In
te
rc
o
n
n
e
c
tio
n
Coaxial Connection
Connect the transducer to the Milltronics transceiver via a junction box and
RG–62 A/U coaxial cable. This setup is effective for combined runs up to
365m (1200’).
Note:
When connecting to an EnviroRanger ERS 500, do NOT use coaxial
cable, see diagram below for proper procedure.
2-Wire Extension
(for EnviroRanger ERS 500 only)
Maintenance
Normally, no maintenance is required on the transducer.
However, if performance changes are observed, immediately shut down the
level measurement system and perform a thorough inspection, especially on
the transducer.
extend cable using
RG – 62 A/U coax
blk
drain / shield
wht
extend cable using 18 AWG
shielded / twisted pair
junction box
Содержание XRS-5
Страница 1: ...33455900 Rev 2 0 XRS 5 XRS 5CTRANSDUCER XRS 5 TRANSDUCER Instruction Manual PL 590 January 2001 ...
Страница 4: ...Page 4 XRS 5 Transducer PL 590 ...
Страница 6: ...Page 6 XRS 5 Transducer PL 590 Specifications ...
Страница 10: ...Page 10 XRS 5 Transducer PL 590 Installation ...
Страница 15: ...PL 590 XRS 5 Transducer Page 15 Applications Water Wastewater Differential Level Pump Control Sewage Lift ...
Страница 18: ...Page 18 XRS 5 Transducer PL 590 Applications ...
Страница 21: ...PL 590 XRS 5 Transducer Page 21 Installation Diagram Installation Diagram for XRS 5 ...
Страница 22: ...Page 22 XRS 5 Transducer PL 590 Installation Diagram Installation Diagram for XRS 5C ...
Страница 23: ... 7ml19981bj01 ...