Maintenance
35
I
NSTALLING THE NEW
P
ROGARD
TL
Remove the new Progard TL from its shipping box. Verify the Millipore SAS Catalogue Number by looking at the sticker on
the side of the Progard TL. Remove the plastic covers on the 4 ports of the Progard TL Pack. Look inside the ports to locate
the black rubber O-rings. Make sure these are pushed in so they are firmly in-place. Wet the Progard TL Pack O-rings with
water.
Look below the ports on the bottom of the Progard TL. Locate the TAG Sensor.
After the used Progard TL is removed, install the new Progard TL with the TAG Sensor at the bottom. The Elix System will
automatically identify the type of Progard TL by accessing data contained on the Pack TAG. Normally the Elix System will
recognise and accept the information from the Progard TL.
Make sure the top and bottom ports of the new Progard TL are pushed in. After this, secure it in place by pushing down on
the Pack Locking Handle.
The Elix System will now rinse the new Progard TL. This is done to empty it of air and to hydrate the material inside.
If you have a Storage Reservoir with a Vent Filter, then the LCD will remind you to change the Vent Filter at this time.
If the Progard TL does not have a chlorine tablet inside of it, then the Elix System will go into SYSTEM CLEANING Mode for 5
minutes.
If the Progard TL does have a chlorine tablet inside of it, then the Elix System will go into SYSTEM CLEANING Mode for 20
minutes.
A DEPRESSURISATION Mode and a PRESSURE TEST Mode will occur for a few seconds prior to the 5 minute or 20 minute
SYSTEM CLEANING Mode.
The CANCEL Keypad Button is not shown if the
Progard TL is brand new and never used.
P R O G A R D T L
M A D E B Y M I L L I P O R E
C O N T I N U E
P L E A S E I N S E R T
N E W . . .
P R O G A R D T L P A C K
C A N C E L
P R O G A R D T L
M A D E B Y M I L L I P O R E
C O N T I N U E
P L E A S E E X C H A N G E
V E N T F I L T E R . . .
C O N T I N U E
P L E A S E E X C H A N G E
V E N T F I L T E R . . .
C O N T I N U E
T E S T I N G F E E D P R E S S .
P L E A S E W A I T . . .
S Y S T E M C L E A N I N G . . .
R . T I M E : 2 0 m i n
S Y S T E M C L E A N I N G . . .
R . T I M E : 5 m i n
T E S T I N G F E E D P R E S S .
P L E A S E W A I T . . .