Maintenance
48
4-9
H
OW TO REPLACE THE
A10
UV
L
AMP
The A10 UV Lamp is normally replaced once per year (every 365 Days).
A message, EXCH. A10 UV IN 14 DAYS, will appear indicating that the A10 UV Lamp needs to be replaced.
It is recommended to have a Millipore SAS Field Service Representative change the A10 UV Lamp. The replacement of the A10
UV involves removing the cover of the system. The instructions for replacing the A10 UV Lamp are not included in this manual.
The instructions are included with the replacement UV Lamp.
Contact Millipore SAS for a copy of these instructions if you would like to see them before ordering a replacement A10 UV Lamp.
4-10
H
OW TO
C
ALIBRATE THE
A10
TOC
M
ONITOR
It is recommended to recalibrate or at least to re-verify the A10 TOC Monitoring Device once a year.
The Elix System
will
not
display any type of message indicating that the A10 needs to be recalibrated.
The A10 TOC Monitoring Device can not be recalibrated on-site. It is necessary to contact Millipore SAS when you need to have
the A10 recalibrated.
Please note that if the A10 is sent to Millipore SAS for recalibration, then the A10 UV Lamp will also be replaced by Millipore SAS
as part of the recalibration procedure.