Millar MPVS-400 Скачать руководство пользователя страница 2

MPVS-400 Quick Start 

M.I. P/N:004-2149 Rev. B 

16.

 

Configure the Chart software according to user preferences (please refer to the ADInstruments user’s 
manual(s) for more details). The PVAN CD contains a suggested Chart data acquisition interface to help 
investigators understand how to utilize the Chart software to record P-V data from the MPVS-400. 

17.

 

Calibration Setup

:  Follow the instructions in the ADInstruments Chart user’s manual(s) as well as Millar’s 

PVAN user’s guide to perform two-point calibration procedures for both the pressure and volume 
channels. For simplified electronic calibration verification, use the transducer/calibration controls on the 
front of the MPVS-400 to provide fixed voltage signal references for pressure (0, 25, 100 mmHg) and 
volume (5, 25, 50 RVU) that may be used to quickly perform the two-point calibration procedure. 
Remember that RVU verification is only for Relative Volume Units, not absolute volume. Conversion of 
Relative Volume Units into true volume units requires additional steps (please refer to Millar’s PVAN 
user’s guide for additional instructions regarding volume calibration techniques). Volume calibration for a 
given catheter may be performed using small tubes of known diameters (a calibration cuvette, which can 
be supplied by Millar Instruments) filled with fresh, slightly heparinized blood drawn from an 
experimental subject. Validation of the volume estimates provided by the Millar hardware can be 
performed by other means, e.g. by thermodilution, flow probes, hypertonic saline dilution, echo, etc. 

18.

 

Pressure Transducer Balancing:

 After the catheter tip has soaked for at least 30 minutes, set the Pressure 

Transducer/Calibration switch on the front panel of the MPVS-400 to Transducer. Unlock and adjust the 
Balance control knob for the pressure channel the catheter is connected to. Use the Balance control knob 
to adjust the pressure transducer output while dipping the transducer just below the surface of sterile 
water or saline (essentially a zero pressure environment). Adjust the Balance control knob until the output 
signal reads approximately 0 mmHg (as displayed by an indicator in the data acquisition software or 
similar monitor) then lock the knob into position. Repeat the procedure if an additional pressure 
transducer is connected to the MPVS-400. Pressure calibration may be verified using an external 
manometer (digital or mercury). 
 
For best stability and zero adjustment, shield the sensor from bright ambient light during the balancing 
procedure and allow the catheter-tip to soak in body temperature sterile water or saline for 30 minutes 
prior to insertion into the biological environment. Zero balancing of the pressure transducer should be 
done prior to each catheterization procedure. 

19.

 

The system is now ready to begin conducting P-V experiments. Prepare the experimental subject for 
catheter insertion according to the lab’s established experimental protocols. When the experimental 
subject is ready, insert the catheter into the left ventricle and begin acquiring and recording P-V data 

[Please Note: Millar Instruments recommends a data acquisition sampling rate of 1 kHz (1000 
samples/second)  in order to accurately capture all of the features of the P-V waveforms 
produced by the fast beating hearts of mice and rats]

. To ensure activation of the catheter tip 

transducers, make sure that both the Pressure Transducer/Calibration Switch and Volume 
Transducer/Relative Volume Units (RVU) Calibration Switch on the front panel of the MPVS-400 are 
switched away from the calibration reference settings and are set to Transducer. Please refer to the 
ADInstruments Chart software guide for additional details on operating the data acquisition software. 
After the P-V data has been collected, please refer to the PVAN software user’s guide for instructions on 
how to export P-V data selections out of the Chart software to perform a comprehensive data analysis in 
PVAN. 

20.

 

When necessary, clean the MPVS-400 hardware with a cloth and isopropyl alcohol or water. Do not allow 
liquids to enter the chassis as this will damage the sensitive electronics housed inside. Air vents may be 
cleaned with suction (i.e. a vacuum cleaner). The MPVS-400 hardware has not been tested for 
compatibility with any method of sterilization. 

www.millarinstruments.com 

Отзывы: