ULTRASONIC PORTABLE MILK ANALYZER
www.milkotester.com
-24-
Operating Instructions Master Pro
Remove the protective cup of the electrode.
Dip the electrode into the low value buffer solution and make sure that
the junction zone of the electrode is at least at 30 mm below the
surface of the buffer solution and leave it at rest. Now you have to
enter the known value of the “low” buffer solution. Use
↑↓
(Up and
Down) arrows to enter the new value and Press Enter to confirm. After
dialing the last symbol press ENTER to confirm the new value.
The following text will appear on the display:
Wait until the electrode and the buffer stabilize /the reference value
should change the least possible/. Press Enter to set the value.
The following text will appear on the display:
Dip the electrode into the high value buffer solution and make sure that
the junction zone of the electrode is at least at 30 mm below the
surface of the buffer solution and leave it at rest. Now you have to
enter the known value of the “high” buffer solution. Use
↑↓
(Up and
Down) arrows to enter the new value and Press Enter to confirm. After
dialing the last symbol press ENTER to confirm the new value.
The following text will appear on the display:
Wait until the electrode and the buffer stabilize /the reference value
should change the least possible/. Press Enter to set the value
Please note:
Setting
Ref. Value: xxxxx
Set
pH high buffer:
x.xx
Setting
Ref. Value: xxxxx