ULTRASONIC PORTABLE MILK ANALYZER
w w w . m i l k o t e s t e r . c o m
CHAPTER SIX
Troubleshuting
Error
message
Possible
problem /cause
Repair/remedy
Empty
Sensor
Insufficient
quantity of the
milk sample
sucked in the
system or air in
the sample
- Check that there are no air bubbles
in
the sample.
- Check if after starting measurement,
milk sample in the sample holder
decreases. In other case – there is
damage in the suction system.
- Check if the end of the suction pipe is
not above the surface of the liquid
(not dipped enough).
- Check the hose connections.
Sample
Overheated
Sucked
overheated
sample
- The sample temperature exceeds the
maximum permissible sample’s
temperature.