MILKOTESTER 

 

Milk analyzing device 

Model – MASTER MINI 

 

Fat, Solids-Non-Fat (SNF), Protein, Lactose, Water content 

Temperature (°C), Freezing point, Solids, Density 

 
 
 
 

 

 
 
 

 

 
 
 

THE INFORMATION CONTAINED IN THIS MANUAL IS A SUBJECT 

TO CHANGE WITHOUT NOTICE. FOR UPDATES, PLEASE 

CONTACT THE MANUFACTURER 

OR USE THE FOLLOWING WEB ADDRESS: 

www.milkotester.com

Содержание MASTER MINI

Страница 1: ...ing device Model MASTER MINI Fat Solids Non Fat SNF Protein Lactose Water content Temperature C Freezing point Solids Density THE INFORMATION CONTAINED IN THIS MANUAL IS A SUBJECT TO CHANGE WITHOUT NO...

Страница 2: ...a leveled and stable surface If it falls or is severely shocked its functional systems may be damaged When plugging the device into the electrical outlet put away the power cord so it does not stay in...

Страница 3: ...Preparing the device 9 4 2 Measuring samples 10 4 3 Making correction and calibration 11 4 3 1 Correction 13 4 3 1 Calibration 14 4 4 Settings and options 18 4 5 Connecting to printer 20 5 Cleaning an...

Страница 4: ...ntioned types of milk The device has a compact design with a robust structure and a user friendly interface MASTER MINI can analyze three types of milk defined by user The measurement speed is 40 45 s...

Страница 5: ...e up view IN THIS CHAPTER Front panel Display Back panel 2 1 Front panel 8 7 9 6 5 1 4 3 2 1 Intake pipette 2 Start knob 3 Operation indicator 4 Arrow DOWN button 5 Arrow UP button 6 ENTER button 7 EX...

Страница 6: ...ULTRASONIC PORTABLE MILK ANALYZER www milkotester com 2 2 Back panel 3 1 2 1 Serial port to printer or PC 2 Power in 12V 3 Power switch 5...

Страница 7: ...es only once and do not return it to the vessel Samples can be kept for a maximum of 2 days if they are stored at a temperature not exceeding 5 Milk stirring It is a very important condition for recei...

Страница 8: ...ermally treated When examining samples taken immediately before analyses and shortly stored the milk is poured several times from vessel to vessel in order to distribute the fat content uniformly To a...

Страница 9: ...d stable surface Caution Any sources of hot or cold air can influence the accuracy of the measurements 2 Connect the device to 12V DC power source 220V AC to 12V DC adaptor or 12V DC electrical lighte...

Страница 10: ...ou can choose among three types of milk at a time When you choose the desired type of milk you use the UP and DOWN arrow buttons and then turn the knob clockwise until the knob indicator comes to the...

Страница 11: ...start measuring anew by turning the knob to the red point By choosing the EXIT button you can start from the very beginning and choose to analyze another type of milk To access the menu press the ENT...

Страница 12: ...reduce this inaccuracy due to water in the system please refer to chapter Cleaning and Maintenance section Thorough Cleaning 4 3 MAKING CORRECTIONS AND CALLIBRATION In the process of work with the ana...

Страница 13: ...ween the values of the parameter from the reference sample and measured with the analyser Make analyses of the difference received If the received differences are relatively constant value for samples...

Страница 14: ...rows and press ENTER to select Correction Calibration Back up or Restore Correction Calibration Back up Restore If you choose correction and press ENTER then use the up and down arrows and press ENTER...

Страница 15: ...4 3 2 Calibration MENU Mode Calibration In order to perform calibration of the device for a certain type of milk for example cow milk You need two samples of milk with known values of their parameters...

Страница 16: ...me procedure to enter the second and third figure After confirmation of the last figure the cursor goes to the next parameter for entering new values mind that there are 2 two pages of components thus...

Страница 17: ...ing the knob counter clockwise to the black point the following message will appear on the display Put Sample High Milk 2 5 and turn to red point Follow the procedure until the 5th measurement After p...

Страница 18: ...end of the 5th measurement the calibration procedure is complete Please note Before each sample measurement stir the milk by pouring it from one vessel to another The temperature of the samples should...

Страница 19: ...tem info Choose Options and press ENTER System Info Language Time and Date Choose System info Language or Time and Date and press ENTER MILKOTESTER Model MASTER MINI Ver x xx SN XXXXXX MENU Options La...

Страница 20: ...wing text appears on the display Set Fr Point 0 520 Press ENTER To Set Use Up and Down arrows to enter the new value and Press Enter to confirm After dialing the last symbol press ENTER to confirm the...

Страница 21: ...to work only with printers supplied by Milkotester Ltd We warn you that the unit is not set to work with other types of printers To connect MILKOTESTER MASTER MINI to the serial printer follow the ste...

Страница 22: ...s or a considerable number of measurements have been performed the device begins to issue continuing beep sounds At the same time the message Cleaning needed appears on the display Please carry out th...

Страница 23: ...to be well stirred up and can remain good for unlimited time After flushing with the solution flush the system with clean water twice by repeating steps 1 and 2 from routine cleaning 5 3 Weekly cleani...

Страница 24: ...e Check that there are no air bubbles in the sample Check if after starting measurement milk sample in the sample holder decreases In other case there is damage in the suction system Check if the end...

Страница 25: ...rotein Lactose Water content Temperature C Freezing point Solids Density 2 Ranges of measurement Fat from 0 00 to 20 SNF from 3 to 15 Density from 1015 to 10 40 kg m3 Proteins from 2 to 7 Lactose from...

Страница 26: ...3 Accuracy Fat 0 1 SNF 0 1 Density 0 3 kg m3 Proteins 0 1 Lactose 0 1 ADDED Water 3 0 Sample Temperature 1o C Freezing point 0 002 o C Solids 0 05 Air temperature from 10 o C to 40 o C Relative humid...

Страница 27: ...st regions is at about 0 540 The average reading for your region is called basic freezing point The freezing point of milk is a physiological constant This does not mean that it will not vary In fact...

Страница 28: ...es obtained as specified in the METHODS After determination the freezing point of your sample via the milk analyzer the added water is calculated using the following formula 100 int int int Base Calc...

Страница 29: ...ct basic freezing point in the device The device s results for added water may give information about doubt of added water in the milk and the exact value of this added water may be determined after a...

Страница 30: ...Botev St 4470 Belovo BULGARIA Milkotester MASTER MINI Guarantee period is 1 one year after purchasing date Improper handling transport and storage will invalidate the guarantee Guarantee is void if w...