Stop a workout immediately if you feel unwell, have muscle
cramps, or are dizzy. Drink fluids and, if problems persist, see a
doctor if necessary.
A basic requirement of EMS training is that the athlete is super-
vised by their trainer. The trainer can therefore intervene appropri-
ately in the event of an emergency. Act immediately if an athlete
loses consciousness:
Personnel:
n
Trainer
1.
Press the multi-function button.
ð
The current training stops.
2.
Turn off the device via the on/off switch on its underside
3.
Detach the i-body
®
connect wireless incl. the cable from the
athlete. See the “i-body
®
connect wireless” manual for further
information.
4.
Initiate first aid measures.
Stop button
The multi-function button is not only used for choosing and con-
firming the programs, but also as a stop button. The training can be
stopped at any time with the multi-function button.
Turning up the intensity during the pulse pause is limited to 5 units
both on the level controllers and the main controller.
Main controller
If the “Load level” function is enabled, the device is programmed
such that the main controller cuts out at 30, 40, and 50 for safety
reasons. Regardless of whether or not the “Load level” function is
enabled, an additional cut-out is set for all programs and training
plans when the device reaches a value of 85. Before the athlete
can increase the intensity beyond this value, the main controller
must be acknowledged by pressing it once.
What to do in emergency situa-
tions?
1
Fig. 56: On/off switch
Stop button
Turning up the intensity
Overview of programs/training plans
2021-07-28
miha-bodytec m.ove EMS training device