Areas of application
19
General information
This washer-disinfector can be fitted with a variety of mobile units,
which can be equipped with a variety of modules and inserts depend-
ing on the type and shape of items requiring cleaning and disinfec-
tion.
Select mobile units, modules and inserts which are appropriate for
the application.
Notes on the individual areas of application and examples of loading
are given on the following pages.
Before starting a
programme
you should carry out a visual check on the following:
– Is everything correctly loaded/connected for cleaning?
– Are the spray arms clean and do they rotate freely?
– Are the filters clean?
Remove any coarse soiling and clean them properly if necessary.
– Is the mobile unit correctly connected to the water supply?
– Are the removable modules, jets, sleeves and other rinsing attach-
ments correctly connected?
– Are all chemical containers sufficiently filled?
At the end of each
programme
The following must be checked at the end of every programme:
– Carry out a visual check of the load for cleanliness.
– Check that all hollow shafted instruments are still securely located
on their jets.
Any hollow instruments that have become disconnected from
their adapters during processing must be re-processed.
– Check that the lumen of hollow instruments are free of obstruction.
– Check that jets and connectors are securely held in position in the
baskets or inserts.
Protein test
Cleaning results should be subjected to periodic protein tests, e.g.
weekly with the Miele test kit or the Miele ProCare Protein Check.