en - Areas of application
38
Reprocessing
Risk of infection due to insufficient disinfection.
Load items that become detached from the irrigation connector
during reprocessing will not be sufficiently disinfected on the in-
side.
Any items that become detached from the irrigation connector dur-
ing reprocessing must be reprocessed again.
Reprocessing
double hoses
used in nitrous
oxide sedation
systems
Make sure that the double hoses can be flushed through.
To check before
loading the ma-
chine and before
starting a pro-
gramme
- Is the silicone insert in the hose adapter in the correct position?
- Are the ends of the double hose inside the basket area?
Содержание APWD 325