en - Wall mounted FlowMeter / ConductivityMeter
44
The water quality display devices can be mounted outside the Aqua
Purificator on the wall.
A suitable wall bracket is supplied with the display devices.
For alternative installation the sequence of individual components
must be observed:
The display device is always connected behind the water deminerali‐
sation cartridge, i.e. in the following order:
water stopcock
- non-return valve with pipe aerator according to
DIN EN 1717
- water demineralisation cartridge
- display device
FM/CM
- washer-disinfector .
Electrical connection
is made via a switched plug and socket.
Maintenance
The PG 8595 Aqua Purificator should be serviced at the same time as
the washer-disinfector by Miele Service.