NECK 

 

 

 

Always place the pads symmetrically. 
 
Treat through the problem area and the largest muscle groups in the region. 
 

Over problem area 

Over middle fibres of Trapezius 

 
Always treat to a maximum comfortable Intensity for the patient. Higher 
Intensity levels are often possible in second and subsequent treatments. 
 
High frequency analgesia/anaesthesia should be used on the channels over 
the pain site. During this procedure, channel B may be operated in the Low 
frequency therapy mode. 

Содержание H-Wave 4-pad

Страница 1: ...Wave 4 pad Instruction Manual A New Concept in electrotherapy MIE Medical Research Ltd Tel 44 0 113 2793 710 6 Wortley Moor Road Fax 44 0 113 2310 820 Leeds e mail hwave mie uk com LS12 4JF United Ki...

Страница 2: ...Timer 7 Battery 8 Warning 8 Re charging the battery 8 Battery Charger Instructions 9 Features 9 Operation 10 Using the H Wave 11 Points to Remember 12 Therapy 13 Notes 14 Pad Placement 15 Neck 16 Shou...

Страница 3: ...onment This Product complies with EN60601 1 2 for EMC and interference potential Use only MIE approved accessories to maintain integrity No user serviceable parts For service contact MIE or your suppl...

Страница 4: ...rainage through an injury site The dramatic increase in blood flow and lymph flow accelerates the removal of pain producing chemicals and toxins which build up around the injury thus reducing pain The...

Страница 5: ...nsity is the strength of the H Wave signal The higher the intensity the stronger and deeper the signal will penetrate The unique signal of the H Wave allows excellent penetration with little skin resi...

Страница 6: ...dependent of each other The control knobs lead sockets and leads are all colour coded to make identification easy Each channel is independently turned on by its intensity control amber light on Place...

Страница 7: ...a red LED display indicates the timer function Initially it will read Co meaning Continuous use To change the time to operate for a fixed period press the black up down buttons underneath the display...

Страница 8: ...that the H Wave is charged over night after use The H Wave charger is specially designed to reduce the charge to a trickle when the battery is fully charged In a low battery situation the battery mus...

Страница 9: ...he internal case temperature rises above a pre determined level the system is designed to automatically shut down the charging current thus reducing the case temperature As the temperature reduces the...

Страница 10: ...een LED on Red LED off Note On power up the charger will appear to be in float mode for a short period of time This is to complete all checks as described in feature soft start Charger Status LED Stat...

Страница 11: ...nt to improve patient comfort For example with a patient suffering from acute low back discomfort a 10 minute session of High Frequency analgesia anaesthesia should control the pain sufficiently for a...

Страница 12: ...tive solution Replace the pad If discomfort still occurs re site the pad slightly away from its original position 6 Always increase the Intensity slowly The response of the patient should be carefully...

Страница 13: ...e Intensity down to 1 and then the Frequency up to high Very gradually increase the Intensity to a high setting and treat for 10 minutes at High Frequency if discomfort occurs it may be advisable to p...

Страница 14: ...d week and subsequently as necessary In acutely painful cases it is often advisable to commence with a few minutes anaesthesia before therapy In these instances the Frequency should be set to High and...

Страница 15: ...er the injury site The pads to Channel B should be placed over the nearest major muscle This ensures maximum movement of fresh blood and lymph fluid through the area during therapy When using High Fre...

Страница 16: ...Over middle fibres of Trapezius Always treat to a maximum comfortable Intensity for the patient Higher Intensity levels are often possible in second and subsequent treatments High frequency analgesia...

Страница 17: ...s of Trapezius and Deltoids Always treat to a maximum comfortable Intensity for the patient Higher Intensity levels are often possible in second and subsequent treatments High frequency analgesia anae...

Страница 18: ...ed B Below elbow A Through the problem area B Over Brachialis Forearm Flexors marked C in diagram Always treat to a maximum comfortable Intensity for the patient Higher Intensity levels are often poss...

Страница 19: ...area B Over Erector Spinae Motor points T10 T12 Always treat to a maximum comfortable Intensity for the patient Higher Intensity levels are often possible in second and subsequent treatments High fre...

Страница 20: ...ctor Spinae Motor points T10 T12 Always treat to a maximum comfortable Intensity for the patient Higher Intensity levels are often possible in second and subsequent treatments High frequency analgesia...

Страница 21: ...ctor Spinae groups Always treat to a maximum comfortable Intensity for the patient Higher Intensity levels are often possible in second and subsequent treatments High frequency analgesia anaesthesia s...

Страница 22: ...ust distal to the head of the Fibula Plantar nerve on the sole of the foot Always treat to a maximum comfortable Intensity for the patient Higher Intensity levels are often possible in second and subs...

Страница 23: ...d B Over Vastus Medialis and Vastus Lateralis Always treat to a maximum comfortable Intensity for the patient Higher Intensity levels are often possible in second and subsequent treatments High freque...

Страница 24: ...ough largest muscle group groups in the area Always treat to a maximum comfortable Intensity for the patient Higher Intensity levels are often possible in second and subsequent treatments High frequen...

Страница 25: ...e have been many cases of successful treatment with the permission and supervision of the physician or surgeon Any such treatment should be discussed with them prior to application In very rare circum...

Страница 26: ...e footnotes to each page of Pad Placement instructions No user serviceable parts For service contact MIE or your supplier Not suitable for use with flammable gases Protect against water Do not use in...

Страница 27: ...elf adhesive pads suitable for H WAVE and purchased from MIE Medical Research Ltd The wires from the pads connect with the colour coded leads by a push fit connection Always ensure that the pad lead c...

Страница 28: ...rode pads Electro conductive solution Battery charger unit Comprehensive User s Guide Under normal conditions the H WAVE can be used continuously This product is a Class 2a medical device and is CE ce...

Отзывы: